MAINTENANCE
4.1
Introduction
This section details the maintenance tasks to be carried out by the operator. The
maintenance chart shows the maintenance tasks, the periodicity of each task and a code
referenced to the task detailed in 4.3.
* Periodicity may change depending on usage and environment, refer to the Engineering Authority.
T
ABLE
4-1 M
AINTENANCE
C
HART
Waste Electrical and Electronic Equipment directive 2002/96/EC
Do not dispose of this product as household waste. Use an approved
organisation that collects and/or recycles waste electrical and electronic
equipment. For more information, contact one of these:
•
our customer service department:
Contact us at www.ge-mcs.com
•
your local government office.
Task
Code
Period
Inspect
A
Daily, before use.
Inspect
B
Weekly
Test
C
Before use.
Test
D
Daily, before use.
Calibrate
E
Every 12 months *
Replace
F
As detailed in fault finding or when detailed after inspection.
Clean
G
Weekly *
Service
H
1,000 hours recorded by the unit's operational timer or the
equipment running log.
Service
J
3,000 hours recorded by the unit's operational timer or 1000 hours
recorded by pump operational timer or the equipment running log.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...