6 - 10
Reference and Specification
K0260 Issue No. 8
[EPR]
•
Engine Pressure Ratio only operates in units of pressure measurement (mbar, inHg etc.) with
a set of pre-defined limits, see section 1. The ADTS 505 can be used to check EPR sensors,
indicators and EPR data for engine control systems.
Note:
Altitude correction applies to EPR values, make sure this correction does not adversely
affect accuracy. The difference in height between the ADTS 505 and the EPR sensor may
be different to the altitude correction previously set for the aircraft instruments.
•
For EPR testing, use Ps as INLET pressure and EPR to set the ratio of OUTLET pressure (Pt) to
INLET pressure (Ps).
•
Before entering a new EPR aim, the rate of change should be set by selecting [EPR Rate] on
the Main Menu.
•
When in Control Mode (AIM display), a new EPR aim can be entered using the numeric keys.
Use
LEAK MEASURE/CONTROL
to change between Control and Leak Measure mode.
Rate EPR
•
When in Control Mode (AIM display), the rate of pressure change can be set using the
numeric keys. Use
LEAK MEASURE/CONTROL
to change between Control and Leak Measure
mode.
[Hold]
•
Selecting [HOLD] causes a changing pressure to instantaneously hold. Selecting [HOLD]
again releases the pressure to continue to the aim value. The [HOLD] facility applies to both
channels simultaneously.
•
[HOLD] automatically releases when
LEAK MEASURE/CONTROL
is pressed entering Measure
Mode.
[Nudge]
•
Changes the current aim value [ or ] by a set increment value (2 ft, 0.2 kts, Mach 0.02).
•
This facility can be used to align a pointer of an instrument to a precise indicated value and
compare this indicated value with the measured value.
•
This facility may be used for pressure switch testing as follows:
¤
Aim for a pressure just below the expected operating point.
¤
When the pressure is achieved, repeatedly press until the switch operates.
¤
Record the pressure reading shown on the display.
•
Repeat this procedure using the until the switch returns to its original condition, refer to
section 3 for an example.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...