8 Troubleshooting
8-15
897N201058J FDC NX700 Instruction Manual 2022.04
(6) When application is poor
Error
code
Error message
Error condition
Troubleshooting
Sample application problem
Because the sample is
not applied to the slide,
the error mark [<] or [@]
appears continuously.
Check the following points.
(a) Make sure that FUJI DRI-CHEM
AUTO TIPS are being used.
(b) Make sure that tips are not being
reused.
(c) Make sure that there are no bub-
bles on the sample surface in the
sample tube.
(d) Make sure that a sample rack that
matches the blood collecting tube
or FUJI tube is being used.
(e) Inspect the sampler O-ring. (Refer
to Section 6.6.1)
IMPORTANT:
Measurement values
with an [@] mark
might be incorrect.
Run the measure-
ment again.
IMPORTANT:
FUJI DRI-CHEM AUTO TIPS cannot
be reused.
NOTE
If errors still occur often, please
contact your dealer for assis-
tance.
8