1-9
1 Safe usage and handling precautions
897N201058J FDC NX700 Instruction Manual 2022.04
1.18 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and FUJI
DRI-CHEM MIXING CUPS
NOTE
In this manual, FUJI DRI-CHEM AUTO TIPS may be shortened to “tip”, FUJI Sample Tube
to “tube”, and FUJI DRI-CHEM MIXING CUPS to “mixing cup”.
FDC NX700 performs sample application automatically via the sampler. However, it is necessary to
prepare dedicated FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and specified blood collecting
tubes and FUJI DRI-CHEM MIXING CUPS.
The FUJI Sample Tubes include the following types.
• FUJI HEPARIN TUBE 1.5 mL
• FUJI HEPARIN TUBE 0.5 mL
• FUJI PLAIN TUBE 1.5 mL
• FUJI PLAIN TUBE 0.5 mL
Do not touch used tips, tubes, and mixing cups with your bare hands as this may cause infection. If
you accidentally touch an item, immediately rinse the contaminated area thoroughly under running
water, and then disinfect the area. Seek medical assistance if necessary.
Used tips, tubes, and mixing cups are infectious industrial waste. Process this waste in compliance
with any applicable regulations in your country, such as by incineration, melting, sterilization, or dis-
infection. Furthermore, when entrusting disposal to another party, request processing from a con-
tractor with a license for special management of industrial waste processing work.
FUJI Sample Tubes: Do not use a deformed tube to avoid causing malfunction of the analyzer.
FUJI DRI-CHEM MIXING CUPS: Do not use a deformed cup to avoid causing malfunction of the
analyzer.
Always use new FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, blood collecting tubes, and
FUJI DRI-CHEM MIXING CUPS for each sample. Do not reuse old ones.
Use only specified products. Otherwise, errors or malfunctions may occur.
When the power is shut off during test processing, such as by a power outage, replenish tips and
replace mixing cups. The analyzer may lose the information of the usage positions of the consuma-
bles and use used mixing cups (holes).
FUJI DRI-CHEM AUTO TIPS: Do not touch the tip-end. It may cause damage to water repellent.
FUJI DRI-CHEM AUTO TIPS: Do not contact the tip-end with other devices. It may cause contami-
nation of water repellent.
WARNING:
WARNING:
CAUTION:
IMPORTANT:
IMPORTANT:
IMPORTANT:
IMPORTANT:
IMPORTANT:
IMPORTANT: