3-20
3 Measurement
897N201058J FDC NX700 Instruction Manual 2022.04
A
電解質
一般比色
図-4.5.2-1
ISE slide
CM slide
4. Load samples and slides.
For details of the sample and slide loading
method, refer to Section 3.3 (3).
NOTE
Place an ISE slide with the arrows fac-
ing up and pointing opposite the sample
label of the cartridge (in this case, “A”)
as shown in the illustration.
NOTE
When you measure an ISE slide and
CM slide at the same time, we recom-
mend loading the ISE slide last in the
cartridge.
5. Close the sample cover, and then lock it
with the sampler cover lock key.
WARNING:
Ensure the sampler cover is locked during sample
processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.
6. At the Measurement-preparation screen,
tap the [START] button to start measure-
ment.
3