6-6
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
6.3 Inspecting and cleaning the transfer bar, incubator, sample
application unit, and ISE unit
Measuring results are affected by stains (from blood samples, etc.) inside of the incubator and ISE
unit. These parts should be checked and cleaned at least once every three months. When a slide
transfer error or a tip ejecting error occurred, clean the sample application unit.
6.3.1 Reference plate level check
Perform [Reference plate level check] before cleaning.
1. Tap the button.
2. Tap the [Normal mode] button.
6