6-22
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
6. Check the results of the [Sampler leak
check].
If the check results are [OK], it is not necessary
to replace the O-ring.
NOTE
The check results can be printed by
tapping the
button.
If the check results are [NG], tap the [Replace
O-ring] button, and then replace the O-ring
according to the following procedure.
NOTE
The check results can be printed by
tapping the
button.
7. Once the check ends, open the sampler
cover, and then remove the sampler leak
check tool mounted on the sampler noz-
zle by hand.