3 Measurement
897N201058J FDC NX700 Instruction Manual 2022.04
(4) Checking the date and time
Check that the date and time displayed on the LCD
are correct.
If they are incorrect, set the date and time using
[Date & Time settings] in Normal mode.
IMPORTANT:
If the date and time are not adjusted correctly, the
analyzer may fail to determine the expiration of the
slides and calibration, and measurement results
might be incorrect.
(5) Loading consumables
Load the consumables (tips, mixing cups, diluent, and reference fluid) in accordance with the
type of measurement.
Refer to Chapter 5
(6) Reading a QC card
QR code
The FDC NX700 can compensate for the differ-
ences of the slide production lots by reading the QC
card in the box of slides.
When [Warming up] or [Ready to test] appears on
the LCD, set the QR code on the left side, and then
insert the QC card into the reader fully.
NOTE
If the QC card is dirty or folded, the read er-
ror may occur.