8 Troubleshooting
8-11
897N201058J FDC NX700 Instruction Manual 2022.04
8.2.3 Slide reading errors
Error
code
Error message
Error condition
Troubleshooting
W0173
Slide identification
ERR
No slide information.
Check the slide.
The slide information
on backside of the slide
could not be read cor-
rectly.
Clean the slide reader, and then rerun
the measurement.
NOTE
If the error occurs again, please
contact your dealer for assis-
tance.
8.2.4 Sampler errors
(1) When tip cannot be mounted
Error
code
Error message
Error condition
Troubleshooting
E0082
Sample tip detection
ERR
No tip detected.
Set tips and rerun the
tests from the begin-
ning.
No tip can be detected
on the sample nozzle
after mounting.
The tip is not loaded correctly. Load
a tip correctly, and then rerun the test
from the beginning.
NOTE
Set tips into all holes of the tip
racks after this error has oc-
curred.
E0083
Reference fluid tip
detection ERR
No tip detected.
Set tips and rerun the
tests from the begin-
ning.
No tip has been de-
tected on the refer-
ence fluid nozzle after
mounting.
E0084
Sample tip clogging
ERR
Rerun the tests from
the beginning.
If the error still occurs,
contact your dealer.
A clog was detected
during mounting of the
sample nozzle tip.
If the error occurs again, please contact
your dealer for assistance.
E0085
Reference fluid tip
clogging ERR
Rerun the tests from
the beginning.
If the error still occurs,
contact your dealer.
A clog was detected
during mounting of the
reference fluid nozzle
tip.
8