8 Troubleshooting
8-34
897N201058J FDC NX700 Instruction Manual 2022.04
Error
code
Error message
Error condition
Troubleshooting
W0042
Incorrect sample rack
WARNING
Correct sample rack
not set in place.
Check the sample
rack.
A re-inspection was
attempted for a meas-
urement that uses the
PF, but a rack other
than a PF sample rack
was loaded.
Load a PF sample rack in which a
sample extracted by PF has been
loaded.
A PF sample rack has
been loaded for a con-
trol measurement.
Load a sample rack other than a PF
sample rack.
W0090
ISE slide direction
ERR
Incorrect loading direc-
tion of ISE slide.
Check the ISE slide.
The ISE slide is loaded
in reverse.
Load the ISE slide in the slide cartridge
in the proper direction, and then meas-
ure again.
W1654
Check sample rack
Unusable sample rack
was set in place.
Check the sample.
Contact your dealer in
case the ERR repeats.
A sample rack that
cannot be used is
loaded.
Use a sample rack intended for this
equipment.
NOTE
If the error occurs again, please
contact your dealer for assis-
tance.
W5400
Expired slide
Slide is expired and
cannot be used.
The validity term of the
slide has expired.
Always use valid slides.
IMPORTANT:
Slide validity is not detected in ISE
tests.
NOTE
Depending on the device set-
tings, this warning may not be
generated.
8