8 Troubleshooting
8-35
897N201058J FDC NX700 Instruction Manual 2022.04
8.2.11 ISE measurement errors
IMPORTANT:
ISE measurement values to which error codes have been appended (E1 - E8) may not be correct.
Perform the following troubleshooting and rerun the measurement.
(1) When a self-test error occurred
Error
code
Error message
Error condition
Troubleshooting
E0530
ISE self test gain ERR
For ISE measurement,
turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.
An error was detected
during a self-test per-
formed before the start
of an ISE measure-
ment.
The analyzer must be restarted for an
ISE measurement. If the error occurs
again, please contact your dealer for
assistance.
Tests other than ISE do not require a
restart.
E0531
ISE self test offset
ERR
For ISE measurement,
turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.
8