3-26
3 Measurement
897N201058J FDC NX700 Instruction Manual 2022.04
CRP
一般比色
図-4.9
CRP
CM slide
2.
Load samples and slides.
For details of the sample and slide loading
method, refer to Section 3.3 (3).
NOTE
When measuring a CRP slide and CM
slide at the same time, stack the CRP
slide last or first in the cartridge.
If a CM
slide is measured before or after a CRP
slide, extra tips are used.
START
STOP
0
3
A
B
D
C
E
MYID001
..0160503010
..0160503009
3.
Tap a sample rack display to display
the sample information registration
screen. Configure the sample type to
[Plasma/Serum].
NOTE
To further set the dilution factor to 3x for
CRP measurements, tap the [Dilution]
button, and then specify 3x.
NOTE
If necessary, enter or choose other
sample information.
4. Close the sample cover, and then lock it
with the sampler cover lock key.
WARNING:
Ensure the sampler cover is locked during sample
processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-
pler cover can be opened.
3