8 Troubleshooting
8-24
897N201058J FDC NX700 Instruction Manual 2022.04
8.2.7 Temperature Controller Errors
Error
code
Error message
Error condition
Troubleshooting
E0565
ISE temperature con-
trol ERR
Problem with ISE.
Contact your dealer.
When the measure-
ment results are ap-
pended with a “+” or “-”
mark or an error listed
to the left occurs, the
temperature inside the
incubator is outside the
specified range.
Please contact your dealer for assis-
tance.
E1500
Temperature control
ERR
Check the ambient
temperature.
Clean the air filter.
Reboot the analyzer.
Perform the following troubleshooting.
(a) Check the room temperature is
within 15 to 32 °C. If the tempera-
ture is out of the range, control the
room temperature.
(b) Switch off the power.
(c) Clean the air filter.
(d) Make sure that thumbscrews of the
incubator are tightened properly
and the incubator cable is connect-
ed to the connector properly.
(e) Make sure that the sample ap-
plication unit cover is attached to
the analyzer properly. Tighten the
thumbscrews securely.
(f) Leave the analyzer for about 10
minutes to allow the interior to cool,
and then switch on power.
NOTE
If the error occurs again, please
contact your dealer for assis-
tance.
E1501,
E1505
Temperature control
ERR
Check the ambient
temperature.
Reboot the analyzer.
E1510
ISE temperature con-
trol ERR
Check the ambient
temperature.
Clean the air filter.
Reboot the analyzer.
E1511,
E1515
Temperature control
ERR
Check the ambient
temperature.
Reboot the analyzer.
E1520,
E1521
Room temperature
WARNING
Check the ambient
temperature.
Reboot the analyzer.
8