4-5
4 Applicable usage methods
897N201058J FDC NX700 Instruction Manual 2022.04
10. If measurement ends, the calibration
measurement results screen appears.
NOTE
Store printed calibration measurement
results during the effective period of the
calibration.
NOTE
The calibration compensated coeffi-
cients are saved to the analyzer auto-
matically.
11. Tap the [OK] button to end calibration.
(Supplement) List display area
The lot numbers (6 digits) and validity terms of read
QC cards (CRP) are displayed in a list. Items are
displayed in the order the QC card was read. Items
within their validity term have their dates displayed
in black. The following items are displayed in red:
the date when the calibration validity term has ex-
pired, [Slide expired] when the validity term of the
slide has expired, and [Unexecuted] when calibra-
tion is not been performed.
4