6-12
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
5
2
3
4
1
Thumbscrew
4. Loosen the 5 screws on the upper sur-
face of the incubator, and then remove
the incubator.
Incubator pressure plate
Cell
5. Moisten gauze or a similar material in
sterilizing ethyl alcohol, strongly wring out
the gauze, and then wipe the 13 pressure
plates on the backside of the incubator.
Dry the pressure plates while the incubator
backside (pressure plate surface) is facing up.
Wipe the incubator cells in the same way.
IMPORTANT:
Do not use a cotton swab. The components of the
cotton swab may negatively affect the measure-
ment values.
IMPORTANT:
Do not touch the surface of the pressure plates
with your bare hands. Otherwise, measurement
values may be negatively affected.
IMPORTANT:
Moisten gauze or a similar material with lukewarm
or cold water, and then wipe off the spotted section.
Thumbscrew
Reference plate
6. Loosen the 2 screws fixing the reference
plate and remove the plate from the ana-
lyzer.
6