5-7
5 Loading consumables
897N201058J FDC NX700 Instruction Manual 2022.04
Diluent display
2. With [Warming up] or [Ready to test] dis-
played on the LCD, make sure that the
analyzer is not applying samples.
3. Tap the diluent display on the LCD.
4. The Diluent loading screen appears.
The sample disk turns, and the consumables
are moved to the front.
CAUTION:
Do not touch the sample disk until it stops.
5. Load the tube containing diluent in ac-
cordance with diluent label of the sample
disk.
IMPORTANT:
Be sure to load diluent in accordance with the
diluent label. Otherwise, the measurement results
may be negatively affected.
5