5-9
5 Loading consumables
897N201058J FDC NX700 Instruction Manual 2022.04
5.5 Loading reference fluid
Reference fluid is necessary for ISE measurements. (Refer to
WARNING:
When handling reference fluid, follow biohazard procedures (wear gloves, a lab coat, and safety
goggles) in accordance with the medical handling rules of your facility. If you accidentally touch an
item, immediately rinse the contaminated body part thoroughly under running water.
WARNING:
Discard reference fluid according to the medical handling rules of your facility and related laws.
WARNING:
When the analyzer is applying samples, tap the [PAUSE] button to interrupt sample application (refer
to Section 4.2
), and then load the reference fluid. It is unsafe to set tips during sample application
due to the moving parts of the sampler and the sample disk.
IMPORTANT:
Use a FUJI PLAIN TUBE (0.5 or 1.5 mL) for reference fluid. Otherwise, measurement errors or
malfunctions may occur.
Configure the type of container to be used via [Tube setting for diluent & reference fluid] in Mode
function for administrator (Refer to Section 9.3.15).
IMPORTANT:
Replace FUJI PLAIN TUBES (1.5 mL) within 24 hours.
Replace FUJI PLAIN TUBES (0.5 mL) for each measurement.
0.5
0.3
0.1
FUJI PLAIN TUBE
(1.5 ml)
Pour reference fluid
within the range.
FUJI PLAIN TUBE
(0.5 ml)
0.5
1.0
1. Prepare reference fluid.
Pour the following amounts into each type of
FUJI PLAIN TUBE.
For 1.5 mL tube: 1.0-1.5 mL
For 0.5 mL tube: 0.3-0.5 mL
NOTE
Use reference fluid after it reaches room
temperature.
5