3.7 CRP Tests
NOTE
Each time a new box of CRP slides is used, a new QC card must be read.
NOTE
Calibration is necessary for CRP measurements under the following conditions.
• When the slide lot has changed
• When the validity term of calibration has expired (For the validity term, refer to the “In-
structions for Use” of the slides)
NOTE
If the validity term has expired, a “#” warning mark is printed along with the measurement
results.
NOTE
Specified diluent and mixing cups are necessary for CRP measurements.
NOTE
When the results of a measurement will exceed the measurement range, dilution can be
performed automatically at 21x or more by configuring the dilution factor at the sample in
-
formation registration screen. (When 2x has been set at the sample information registration
screen, the dilution factor is 42x.)
1. Check the remaining volume of the con-
sumables (tips, mixing cups, and dilu-
ent).
If consumables are insufficient, supplement or
replace them.
• Loading tips
• Loading mixing cups
• Loading diluent
NOTE
When performing measurements that
require dilution, at least 2 tips are used.
NOTE
When performing CRP tests that require
dilution at 2x or more, 2 wells (holes) of
mixing cups are used.
Diluent
NOTE
For CRP, set the dedicated diluent in the
determined position.