3-3
3 Measurement
897N201058J FDC NX700 Instruction Manual 2022.04
3.2 Measurement preparation
(1) Cleaning the disposal box
Empty the disposal box.
IMPORTANT:
If the disposal box is full, a slide transfer error or tip
disposal error may occur, which may negatively af-
fect the measurement values.
IMPORTANT:
After emptying the disposal box, be sure to load the
disposal box. If the analyzer is used without the box,
the measurement results may be negatively affected.
WARNING:
Do not touch used slides and tips with your bare
hands, as this may cause infection. If you acciden-
tally touch an item, immediately rinse the contami-
nated area thoroughly under running water, and
then disinfect the area. Seek medical assistance if
necessary.
WARNING:
Used slides and tips are infectious industrial
waste. Process this waste in compliance with any
applicable regulations in your country, such as by
incineration, melting, sterilization, or disinfection.
3