background image

4-25

4 Applicable usage methods

897N201058J FDC NX700 Instruction Manual 2022.04

6.  Close the sample cover, and then lock it 

with the sampler cover lock key. 

WARNING:

Ensure the sampler cover is locked during sample 

processing to prevent injuries and infection.
When the display reads [Ready to test], the sam-

pler cover can be opened.

7.  Tap the [START] button to start measure-

ment.

8.  After measurement is ended, discard the 

used PF, and then remove the sample 

container.

WARNING:

When discarding used PF, always follow biohazard procedures (wear gloves, a lab coat, and safe-

ty goggles) in accordance with the sample handling rules of your facility. If any part of the body 

comes in contact with a sample, immediately rinse the contaminated body part thoroughly under 

running water and then use ethyl alcohol as a disinfectant. Seek medical assistance if necessary.

WARNING:

Used PF is infectious industrial waste. Process this waste in compliance with any applicable regu-

lations in your country, such as by incineration, melting, sterilization, or disinfection.

4

Summary of Contents for FUJI DRI-CHEM NX700

Page 1: ...CHEM NX700 FUJI DRI CHEM NX700i 7th Edition 897N201058J April 2022 INSTRUCTION MANUAL FUJI DRI CHEM CLINICAL CHEMISTRY ANALYZER EN This Manual describes details on how to operate the FUJI DRI CHEM NX700 NX700i and cautions to be observed when operating it Please read this manual thoroughly before actually operating the FUJI DRI CHEM NX700 NX700i After reading this manual store it nearby the FUJI D...

Page 2: ......

Page 3: ...cts of other manufacturers not supplied by FUJIFILM 5 FUJIFILM shall not be liable for malfunctions and damage resulting from remodeling maintenance and repair using repair parts other than those specified by FUJIFILM 6 FUJIFILM shall not be liable for malfunctions and damage resulting from negligence of precautions and operating methods contained in this manual 7 FUJIFILM shall not be liable for ...

Page 4: ...with all the series FUJI DRI CHEM NX700 and FUJI DRI CHEM NX700i the former includes the PF Plasma Filter unit and the latter does not So for the latter sections that describe the unit can be skipped Name PF Unit FUJI DRI CHEM NX700 4 FUJI DRI CHEM NX700i NOTE 4 The PF Unit is mounted on the device The PF Unit is not mounted on the device ...

Page 5: ... 1 17 FUJIFILM PF plasma filters 1 8 1 18 FUJI DRI CHEM AUTO TIPS FUJI Sample Tubes and FUJI DRI CHEM MIXING CUPS 1 9 1 19 Light source lamp 1 10 1 20 Recording paper 1 10 1 21 Labels 1 11 1 22 Symbols 1 12 2 Part names and consumables 2 1 Part names 2 1 2 2 Touch panel screen names and functions 2 5 2 2 1 Measurement preparation screen 2 5 2 2 2 Help screen 2 7 2 3 Names and functions of software...

Page 6: ... 31 4 8 Batch change of sample types 4 35 4 9 Measuring plasma and urine samples at the same time 4 36 4 10 Editing sample numbers and sample IDs 4 38 4 11 Using the sample barcode reader sold separately 4 39 5 Loading consumables 5 1 Loading consumables 5 1 5 2 Setting tips 5 2 5 3 Loading mixing cups 5 4 5 4 Loading diluent 5 6 5 5 Loading reference fluid 5 9 6 Maintenance 6 1 Periodic maintenan...

Page 7: ...eading errors 8 11 8 2 4 Sampler errors 8 11 8 2 5 Photometric system errors 8 20 8 2 6 Transfer Errors 8 22 8 2 7 Temperature Controller Errors 8 24 8 2 8 Circuit board signal errors 8 26 8 2 9 QC card PF card and DI card errors 8 31 8 2 10 Slide loading errors 8 33 8 2 11 ISE measurement errors 8 35 8 2 12 PF errors 8 37 8 2 13 Errors Related to Data Communication Sample Barcode Reader 8 39 8 2 ...

Page 8: ... fluid setting 9 37 9 3 18 Control measurement setting 9 38 9 3 19 Operator ID setting 9 38 9 3 20 Test item name setting 9 39 9 3 21 Password setting 9 40 10 Other functions 10 1 Data communication 10 1 11 USM User Safety Mode 11 1 Operator ID input when measurement start and in mode function 11 1 11 2 Difference of operable functions among operator ID level OPE ADM EMG 11 2 11 3 USM menu 11 5 11...

Page 9: ...llowed NOTE Note indicates procedures requiring special attention instructions that must be followed supplementary explanations etc 1 2 Operation precautions Before using this equipment please read this Instruction Manual carefully so that you can operate the equipment correctly Whenever you operate this equipment be sure to observe the precautions described in this manual Failure to do so may sub...

Page 10: ...tor because it cannot be disposed of as a general waste Not doing so may cause environmental pollution When handling samples blood or urine and performing maintenance cleaning the analyzer al ways follow biohazard procedures wear gloves a lab coat and safety goggles in accordance with the sample handling rules of your facility If you accidentally touch a sample immediately rinse the contaminated a...

Page 11: ...ompatible electromagnetic environment for the equipment can be main tained in order that the device will perform as intended This equipment conforms to the following emission and immunity requirements IEC 61326 2 6 2012 Class B EN 61326 2 6 2013 Class B IEC 61326 2 6 2005 Class B EN 61326 2 6 2006 Class B The electromagnetic environment should be evaluated prior to operation of the device Do not u...

Page 12: ...d to direct sunlight Places near sources of heat such as heaters Places where the temperature may drastically change Places where the equipment is subject to vibration or its support table is unstable 1 Install the equipment in the following environmental conditions Installation environment Indoor use do not expose to direct sunlight Illumination 6 000 cd m2 lux or less 3 000 cd m2 lux or less whe...

Page 13: ...n the device Contact your dealer Failure to observe these precautions may result in unexpected errors or equipment malfunctions When wax was used to clean the floor of the installation location make sure that wax will not adhere to slides and the device If wax has adhered the measurement results may be affected 1 9 Action for Malfunction When the equipment generates unusual noise smell or smoke im...

Page 14: ...e Process this waste in compliance with any applicable regulations in your country such as by incineration melting sterilization or disinfection Further more when entrusting disposal to another party request processing from a contractor with a license for special management of industrial waste processing work Please read the Instructions for Use of the slides carefully and follow the instructions ...

Page 15: ... has been used up NOTE When QC information exceeds 5 lots the oldest QC information is deleted Slides with out QC information cannot be measured 2 A QC card contains the slide test name and the correction coefficient corresponding to the man ufacturing number lot Always use slides from the same box 1 14 Handling samples When handling samples blood or urine always follow biohazard procedures wear g...

Page 16: ...ear gloves a lab coat and safety goggles If you accidentally touch an item immediately rinse the contaminated body part thoroughly under running water Seek medical assistance if necessary For diluents used in measurements that require dilution refer to the Instructions for Use of the slides Use the reference fluid included with the ISE slides for ISE measurements Refer to the Instructions for Use ...

Page 17: ...ste in compliance with any applicable regulations in your country such as by incineration melting sterilization or dis infection Furthermore when entrusting disposal to another party request processing from a con tractor with a license for special management of industrial waste processing work FUJI Sample Tubes Do not use a deformed tube to avoid causing malfunction of the analyzer FUJI DRI CHEM M...

Page 18: ...the lamp turn the power off wait at least 5 minutes and then check that the lamp has cooled 1 The light source lamp is a precision halogen lamp Do not touch the glass surface of the lamp with your bare hands 2 The lamp is a consumable with a lifetime Always keep a spare lamp on hand in case the lamp burns out 1 20 Recording paper 1 Use specified recording paper for the FDC NX700 2 If other recordi...

Page 19: ...X700 Power switch Depress to power ON O Depress to power OFF Moving parts caution label Biohazard label Biohazard label PF tube caution label Refer to Section 6 9 4 Biohazard label High temperature caution label Refer to Section 6 5 Lamp caution label Refer to Section 6 5 Name Plate Sampler cover lock key caution label Refer to Section 1 15 Power button 1 ...

Page 20: ...lt documents Biological risks High temperature caution Power button Power switch Fragile item handle carefully Protect from rain This way up 2 Stacking limitation Temperature limitation Manufacturer SN Serial number Country of manufacture Date of manufacture EC REP Authorized representative in the European Community In vitro diagnostic medical devices 1 ...

Page 21: ...Safe usage and handling precautions 897N201058J FDC NX700 Instruction Manual 2022 04 Sign Description CE mark of conformity Importer Consult an electronic Instructions for Use e IFU Unique device identifier 1 ...

Page 22: ......

Page 23: ...e following is an illustration of the FDC NX700 full option model FDC NX700i has slight difference Printer Card reader LCD touch panel Power button Disposal box Sampler cover Front bottom cover Sampler cover lock Incubator cable Incubator cover Sampler nozzle cover Plasma filter unit PF unit Sample disk 2 ...

Page 24: ...2 2 2 Part names and consumables 897N201058J FDC NX700 Instruction Manual 2022 04 USB connector Power switch Power inlet Air filter Louver LAN terminal RS 232C connector Air filter Louver 2 ...

Page 25: ...es 897N201058J FDC NX700 Instruction Manual 2022 04 Light source lamp Disposal box Sample rack A E Sample application unit cover Slide cartridge A E Reference fluid holder Calibrator holder Mixing cup Tip rack Diluent holder D1 D3 2 ...

Page 26: ...2 4 2 Part names and consumables 897N201058J FDC NX700 Instruction Manual 2022 04 Details of PF unit PF unit PF sensor suction pad PF reference plate PF sample rack PF pump 2 ...

Page 27: ...order information from the host PC via a DMS connection è Refer to Section 4 7 3 Function menu button Display the function menu screen è Refer to Chapter 9 4 START button Start a measurement 5 Operator ID button displayed ac cording to settings If tapped you can enter the operator ID è Refer to Section 9 3 19 6 Feed button Feed the recording paper with a fixed amount PAUSE button displayed during ...

Page 28: ... Section 5 4 0 Reference fluid display If this display is tapped the reference fluid can be loaded è Refer to Section 5 5 a Tips If this display is tapped tips can be loaded è Refer to Section 5 2 b Sample racks A E If a sample rack is tapped the sample in formation registration screen of that rack is displayed è Refer to Section 3 3 4 c Plasma filter PF production num ber lot display Displays the...

Page 29: ...201058J FDC NX700 Instruction Manual 2022 04 2 2 2 Help screen During measurement maintenance consum able exchange and other actions on the FDC NX700 you can tap the button on the Meas urement preparation screen to check the proce dures 2 ...

Page 30: ...2 8 2 Part names and consumables 897N201058J FDC NX700 Instruction Manual 2022 04 2 1 1 button The Measurement preparation screen ap pears 2 button Return to the previous screen 2 ...

Page 31: ...bols and katakana 1 2 5 4 7 6 3 1 Alphabetical characters numbers symbols and katakana buttons Enter alphabetical characters numbers symbols and katakana 2 button Delete all entered characters and then quit keyboard entry 3 Move cursor buttons Move the position of the cursor left or right by one character 4 Clear button Delete all entered characters 5 OK button Confirm the entered characters 6 but...

Page 32: ...n is selected When the button is selected When the ABC button is selected a 0 9 8 8 ABC button Display the entry screen for uppercase alpha betical characters and numbers 9 abc button Display the entry screen for lowercase alpha betical characters and numbers 0 123 button Display the entry screen for numbers com ma and minus symbol a button Display the entry screen for symbols and numbers 2 ...

Page 33: ...TE An AC power cable is not supplied as a bundled item Please prepare an AC power cable conforming to the specifications below The requirements of the cable specifications for the FDC NX700 Power voltage 100V 120V Requirements for the plug connector AC125V 10A Requirements for the cable SVT 3 18AWG 60 C Power voltage 200V 240V Requirements for the plug connector AC250V 10A Requirements for the cab...

Page 34: ...IN TUBE 0 5 mL 500 1 box Recording paper 6 rolls 1 box Light source lamp 1 O ring 4 Plasma Filter PF 1 50 1 bag PF Suction Pad 1 PF Tube 1 FUJI DRI CHEM MIXING CUPS 1 50 1 box Reference fluid cap packing 5 1 bag NOTE Parts names marked with are the same parts packed with the FDC NX700 NOTE 1 Accessories Extra facilities to the analyzer which easily can be mounted by the user Chemicals FUJI DRI CHE...

Page 35: ... weight 1 Sample barcode reader 1 FUJI DRI CHEM CLEAN TIPS 1 for manual sampling 6 cases CAUTION The sample barcode reader specified for the FDC NX700 can be used Do not connect a barcode reader other than specified for the FDC NX700 Otherwise physical damage or danger of fire may result NOTE Parts names marked with are the same parts packed with the FDC NX700 NOTE 1 Accessory Extra facility to th...

Page 36: ...ts Diluent Mixing cup 2 locations Reference fluid Tips 25 2 locations Refer to Section 3 2 1 Refer to Section 5 2 Refer to Section 5 5 Refer to Section 5 3 Refer to Section 5 4 Refer to Chapter 5 Loading consumables Checking the recording paper Cleaning the disposal box Starting the FDC NX700 Checking the date and time DL Step 1 Measurement preparation Refer to Section 3 2 2 Refer to Section 3 2 3 ...

Page 37: ... Refer to Section 3 7 Slide set order CRP 一般比色 図 CRP CM Slide 電解質 一般比色 図 ISE Slide CM Slide CRP 一般比色 図 CRP CM Slide 5倍希釈 2倍希釈 図 4 5 2 2 x5 Dil x2 Dil 1 2 3 Slide Sample Start button Checking the tips Basic measurements Refer to Section 3 3 Checking the measurement results Refer to Section 4 6 To test CRP and CM Colorimetric slide To test CM and ISE slide To test CRP CM and ISE slide To test CM sli...

Page 38: ...box If the analyzer is used without the box the measurement results may be negatively affected WARNING Do not touch used slides and tips with your bare hands as this may cause infection If you acciden tally touch an item immediately rinse the contami nated area thoroughly under running water and then disinfect the area Seek medical assistance if necessary WARNING Used slides and tips are infectiou...

Page 39: ...in the illustration to release the lock 1 and then lift up the cover 2 Recording paper Check the remaining recording paper FUJI thermal paper If the recording paper is short for the meas urement replace the old roll with a new one è Refer to Section 6 4 Cutter opening Recording paper NOTE Before closing the cover pull out the edge of the recording paper 3 ...

Page 40: ...s been shut off during measurement due to a power outage or other issue replenish the tips refer to Section 5 2 and replace the mix ing cups refer to Section 5 3 The analyzer may lose the information of the usage positions of the consumables and use wells holes of used mixing cups Power switch 0013 NOTE When the power switch is not in the ON po sition push the switch to the side NOTE The listing t...

Page 41: ...s and calibration and measurement results might be incorrect 5 Loading consumables Load the consumables tips mixing cups diluent and reference fluid in accordance with the type of measurement è Refer to Chapter 5 6 Reading a QC card QR code The FDC NX700 can compensate for the differ ences of the slide production lots by reading the QC card in the box of slides When Warming up or Ready to test app...

Page 42: ... the LCD When Ready to test appears on the LCD make sure that the analyzer is not applying samples NOTE You can stop sample application by press ing the PAUSE button è Refer to Section 4 2 NOTE During sample application a new sample for the next measurement cannot be loaded on the analyzer 2 Setting tips Tips tip rack Tips Open the sampler cover and then check the tip racks have been loaded with t...

Page 43: ...hem to reach room tempera ture for 5 minutes or more with the condition of the individual package before unwrapping and using them Otherwise the measurement results and the analyzer operation may be affected IMPORTANT After the package has been unwrapped finish measurement within 30 minutes 0043 A Slide weight Slide 2 Remove the slide cartridge of the sample position to be measured from the sam pl...

Page 44: ...that contains sample and then load it on the dedicat ed sample rack NOTE Make sure there are no bubbles in the sample in the blood collecting tube or FUJI Sample Tube Use a centrifuge to remove bubbles If bubbles are present the sample may be detected incorrectly IMPORTANT Be sure not to mistake the positions of the sample and its measurement slides A B C D E Sample rack Slide cartridge NOTE When ...

Page 45: ...the list 6 Select a dilution factor from the list è Refer to Section 3 6 and Section 9 3 14 7 Enter the registration information and then re turn to the Measurement preparation screen 8 Cancel entry of the registration information and then return to the Measurement preparation screen NOTE When the analyzer is connected to DMS sample information can be acquired automati cally from the work list è R...

Page 46: ...o test the sam pler cover can be opened Tap the START button to start measurement IMPORTANT Close the disposal box If it is open the measure ment results will be affected NOTE When starting measurements the analyzer recognizes the presence of sample racks and slides and then measures properly loaded samples NOTE When the measurement conditions have been configured for each sample when measurement ...

Page 47: ...ck dis play blinks during sample application Time until the end of measurement Measurement start time If sample application ends the sample rack display switches off When slides and samples have been loaded in multiple sample racks the next sample rack display blinks orange and then sample applica tion is started 3 ...

Page 48: ...loaded in the cartridge To retest Section 4 5 tap the retest button NOTE For the details of the measurement results screen refer to Section 3 4 7 Ending measurement To complete the measurement unlock the sampler cover and open it and then remove the measured samples NOTE When a measurement is not being performed empty the disposal box as appropriate The disposal box can hold 150 slides and 75 AUTO...

Page 49: ...ood samples start to get separated PF filtering may not occur normal ly and an error may occur NOTE With dilution measurements refer to Section 3 6 you can schedule the start of measurement as follows tap the sample rack display to display the sample information registration screen After configuring the dilution factor in the Dilution field tap the START Reserve button on the Measurement preparati...

Page 50: ...3 15 3 Measurement 897N201058J FDC NX700 Instruction Manual 2022 04 If measurement is possible it starts automatically NOTE To cancel a scheduled measurement start tap the STOP button 3 ...

Page 51: ...l Green Within reference interval Red Outside reference interval 2 3 Measurement results that include a value out side the reference interval are indicated by and the corresponding test name and value become red If tapped the details screen ap pears And will be indicated when an error has oc curred NOTE If measurement items that display H L or other characters are tapped the details screen appears...

Page 52: ...amount 4 Print button Print all measurement results of the dis played sample 5 Send button Send the measurement results of the dis played sample to a PC via DMS 6 Retest button Perform the measurement again under the same conditions for sample number sample ID reference standard and sample type Only the dilution factor can be changed è Refer to Section 4 5 7 Result data scroll button Change the di...

Page 53: ...splay 1 2 1 Measurement month scroll buttons Change the display of the calendar month 2 Date buttons Display the measurement results screen of a date The currently displayed date is dis played in blue NOTE Dates with no measurements are displayed in gray and cannot be se lected 3 ...

Page 54: ...uid Reference fluid display 1 Tap the reference fluid display on the LCD to display the Replacement screen and then load the reference fluid When the loading is complete press the Finish button to close the Replacement screen For the reference fluid container FUJI plain tube 0 5 mL or 1 5 mL use the container con figured for Tube setting for diluent reference fluid in Mode function for administrat...

Page 55: ... this case A as shown in the illustration NOTE When you measure an ISE slide and CM slide at the same time we recom mend loading the ISE slide last in the cartridge è Refer to Section 4 6 5 Close the sample cover and then lock it with the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test...

Page 56: ...TE Do not dilute whole blood samples NOTE The default dilution factors designated by the manufacturer are as follows ISE urine tests 2x CRP tests 21x Excluding ISE tests the following section explains the relationships among the dilution fac tor configured for each sample at the sample information registration screen the dilution factor for each test configured by Dilution setting in Mode function...

Page 57: ...ficient supplement or replace them Loading tips è Refer to Section 5 2 Loading mixing cups è Refer to Section 5 3 Loading diluent è Refer to Section 5 4 2 Load samples and slides For details of the sample and slide loading method refer to Section 3 3 3 START STOP 0 3 A B D C E MYID001 0160503010 0160503009 3 Tap sample rack display A In the Sample information registration screen tap the Dilution b...

Page 58: ...0 0160503009 4 In the list tap x2 Dilution is performed according to Dilution setting in Mode function for administra tor None No dilution NOTE If necessary enter or choose other sample information START STOP 0 3 A B D C E MYID001 0160503010 0160503009 5 Tap the OK button to finish configura tion of Sample A 3 ...

Page 59: ...ock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can be opened 7 At the Measurement preparation screen tap the START button to start measure ment Dilution is performed at the dilution factor speci fied for each sample 3 ...

Page 60: ...sults of a measurement will exceed the measurement range dilution can be performed automatically at 21x or more by configuring the dilution factor at the sample in formation registration screen When 2x has been set at the sample information registration screen the dilution factor is 42x 1 Check the remaining volume of the con sumables tips mixing cups and dilu ent If consumables are insufficient s...

Page 61: ...B D C E MYID001 0160503010 0160503009 3 Tap a sample rack display to display the sample information registration screen Configure the sample type to Plasma Serum NOTE To further set the dilution factor to 3x for CRP measurements tap the Dilution button and then specify 3x è Refer to Section 3 6 NOTE If necessary enter or choose other sample information 4 Close the sample cover and then lock it wit...

Page 62: ...3 27 3 Measurement 897N201058J FDC NX700 Instruction Manual 2022 04 5 At the Measurement preparation screen tap the START button to start measure ment 3 ...

Page 63: ... fluid from the analyzer After removing the items tap the Finish button to rotate the sample disk again Sampler cover 3 Close the sampler cover right side 4 Make sure that measurement and mode operation are not in progress CAUTION Do not switch off the power during measurement or mode operation Otherwise the analyzer may mal function If the power was shut off during measure ment load tips and othe...

Page 64: ...anual 2022 04 Power button 5 Press the power button 6 Tap OK CAUTION Were you not able to operate on the screen press the power button at front to turn off the power NOTE When the analyzer will not be used for a long time set the power switch to the side 3 ...

Page 65: ... water and then disinfect the area Seek medical assistance if necessary WARNING Used slides and tips are infectious industrial waste Process this waste in compliance with any applicable regulations in your country such as by incineration melting sterilization or disinfection NOTE When disinfecting the disposal box use ethyl alcohol for disinfection or a 0 5 sodium hypochlorite solution When a sodi...

Page 66: ...cting tubes that can be used Outer diameter of Ø16 indicates Ø15 0 to 16 6 mm and Ø13 indicates Ø12 0 to 13 3 mm The following illustrations show the design suction ranges If the sample volume is low use a FUJI Sample Tube 1 5 mL or 0 5 mL 測定不能範囲 1 検体は φ16 100mmでおよそ3 5mI以上 φ13 100mmでおよそ2 5ml以上必要です 2 検体は およそ0 5ml以上必要です 10mm φ16またはφ13 x100mm採血管 φ13 x75mm採血管 測定不能範囲 100mm 10mm 30mm 測定不能範囲 2 9mm 測定不能範囲...

Page 67: ...ut general blood collecting tube contents and the inspection items that can be measured Heparin blood collecting tubes Inspection items Applicable to all items For GLU and NH3 measure the samples as soon as possible after blood collection General color of cap Green in Japanese market Plain coagulant included blood collecting tubes Inspection items All other than GLU and NH3 General color of cap Re...

Page 68: ...that the sample volume is sufficient by referring to the illustration to the left The top sur face of the sample fluid must be within the shaded section of the tubes in the illustration NOTE If the sample in the tube is below the suc tion limit mark the analyzer will indicate an insufficient sample volume error NOTE To aspire only the supernatant fluid plasma of the centrifuged sample and avoid as...

Page 69: ...n Manual 2022 04 3 9 3 Sample racks A sample rack is specified for each sample tube Place each sample tube into the specified sample rack NOTE The sample racks are optional 0 5ml 0 5ml 1 5ml 13 75 13 100 16 100 75 13 PF 100 13 PF 100 16 PF Sample rack 3 ...

Page 70: ... compensate for varia tions in X of each lot 2 Inputting lot compensation coefficients using a QC card The lot compensation coefficients are recorded on QC cards and one QC card is packed with each box of slides When data is input from a new card the previously stored lot compensation coefficients are updated Always read the new QC card when you switch to a new box of slides because each QC card a...

Page 71: ...er lot are displayed on the LCD and output to the printer as well NOTE Do not insert another QC card while the slide test and lot number are displayed NOTE When the analyzer beeps and the test is not displayed insert the QC card into the card reader again 4 Using DI cards When you receive DI cards from FUJIFILM due to a change of the slide s manufacturing pro cess read the DI cards according to th...

Page 72: ......

Page 73: ...infect the area Seek medical assistance if necessary IMPORTANT Perform the calibration measurement using Unit A mg dL IMPORTANT When the calibrator cap is opened there may be a liquid film on the bottle opening In this case always remove the film before the calibration process IMPORTANT The cap of the calibrator should not be open more than 5 minutes during a single calibration IMPORTANT Use a cal...

Page 74: ...NX700 Instruction Manual 2022 04 1 Read the QC card via the analyzer Read the QC card of the calibration slide è Refer to Section 3 10 2 Tap the button 3 Tap the Calibration button 4 Tap the Calibration button The Calibration Lot No Screen appears 4 ...

Page 75: ...rator concentration settings refer to Section 4 1 2 6 Tap the START Calibration button CP 1 Blue CP 2 Green CP 3 Red 7 Follow the on screen instructions to load consumables such as slides calibrator or diluent NOTE The following consumables are used by CRP calibration diluent tips and mix ing cups For details of loading each consumable refer to Chapter 5 Load ing consumables NOTE The CRP calibrati...

Page 76: ...truction Manual 2022 04 9 Tap the START button to start calibra tion measurement If a slide with a different lot number has been loaded an error screen appears Tap the OK button to return to the previous screen Check the slide and then continue measurement 4 ...

Page 77: ...alyzer auto matically 11 Tap the OK button to end calibration Supplement List display area The lot numbers 6 digits and validity terms of read QC cards CRP are displayed in a list Items are displayed in the order the QC card was read Items within their validity term have their dates displayed in black The following items are displayed in red the date when the calibration validity term has ex pired...

Page 78: ...the concentration in the calibrator concentration display area at the Calibration Lot No screen are different 1 Tap the Settings button The administrator password entry screen ap pears 2 Enter the administrator password and then tap the OK button The Calibration concentration configuration screen appears NOTE The administrator password is neces sary when moving from the Function menu screen to the...

Page 79: ...Instruction Manual 2022 04 3 Tap the Input field of each CP The software keyboard screen appears 4 Enter a value and then tap the OK button For the details of the software keyboard screen refer to Section 2 3 IMPORTANT Enter the value using Unit A mg dL 4 ...

Page 80: ...nual 2022 04 5 If the button is tapped the entered values are registered and the analyzer returns to the Calibration Lot No screen NOTE If the Reset button is tapped the en tered values are reset If the button is tapped the entered values are regis tered and printed 4 ...

Page 81: ...22 04 4 1 3 Verifying calibration There is a function to verify calibration 1 At the Calibration screen tap the Verifying calibration button 2 Tap the checkbox of a level under Cali bration level selection to select a level to be measured 3 Tap the START Calibration button 4 ...

Page 82: ...e sample cover and then lock it with the sampler cover lock WARNING Ensure the sampler cover is locked by the sampler cover lock during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can be opened 6 Tap the Start button to start verifica tion 7 If measurement ends the Calibration verification result screen appears NOTE If the screen to t...

Page 83: ...you can display the loading screen by touching the display of each consumable You can cancel the remaining sample application by tap ping the PAUSE button in this paused state The measurement of slides with sample applied con tinues 4 2 1 Interrupting and canceling sample application 1 Interrupting sample application Tap the PAUSE button The LCD changes to Spotting is suspended The sample rack dis...

Page 84: ...ed If the cancel operation is completed the LCD changes to Ready to test All sample rack dis plays get switched off 4 2 2 Requesting a STAT test You can request a STAT test by pressing the STAT button NOTE The STAT sample is measured after the sample for which application was interrupted NOTE A STAT test can be requested for 1 sample only Example Requesting a STAT test for Sample C after Sample A ...

Page 85: ...4 13 4 Applicable usage methods 897N201058J FDC NX700 Instruction Manual 2022 04 2 Tap sample rack display C 3 Register the measurement information and then tap the OK button 4 ...

Page 86: ...97N201058J FDC NX700 Instruction Manual 2022 04 4 Tap the RESTART button to re start measurement NOTE In this example the analyzer finishes the remaining application of Sample A and then measures Sample C Measurement order A Þ C Þ B 4 ...

Page 87: ...e area Seek medical assistance if necessary IMPORTANT In the case of a CM slide use a dedicated pipette that matches the sample application volume specifications in the Instructions for Use of the slide Carefully read the operation method of the pipette in Section 4 3 2 Be aware that the sample application volume of GLU W is 6 µL which is different from other slide items NOTE Manual sample applica...

Page 88: ...A comes to the front START STOP 0 3 A B D C E MYID001 0160503010 0160503009 4 Tap sample rack display A In the Sample information registration screen register the measurement information and then tap the OK button 5 Place slides to be tested in Slide cartridge A 6 Aspirate the sample by using the pipette NOTE Refer to the Instructions for Use of the slide for the amount of sample to be aspirated a...

Page 89: ...897N201058J FDC NX700 Instruction Manual 2022 04 7 Tap the START button A slide is transferred to the sample application position 8 Prepare so that sample application with the pipette is always possible and then tap the START button 4 ...

Page 90: ... is sounding while Spot now is displayed Otherwise the measurement results will be inaccurate IMPORTANT When applying sample to a CM slide apply the sam ple within 1 mm of the center of the slide Otherwise the measurement results will be inaccurate Run the measurement again 参照液 検体 搬送方向 イ ンキュベータ方向 スライ ドの中央に点着する 比色スライ ド Apply sample to the center of the slide CM slide NOTE The LCD may not display Wa...

Page 91: ...y rinse the contaminated area thoroughly under running water and then disinfect the area Seek medical assistance if necessary IMPORTANT Use a FUJI clean tip IMPORTANT Mount the tip securely into the nozzle of the pipette If mounting is poor the tip may be dislodged and the sample may scatter IMPORTANT Do not touch the tip end It may cause damage to water repellent IMPORTANT Do not contact the tip ...

Page 92: ... drop 液滴を作ってスライ ドに近づけます 液滴をスライ ドの中央に点着 し ます スライ ドが検体を吸い取り ながら 内部に拡散 していき ます 0057 Place the drop on the center of the slide 液滴を作ってスライ ドに近づけます 液滴をスライ ドの中央に点着 し ます スライ ドが検体を吸い取り ながら 内部に拡散 していき ます 0057 While the slide is absorbing the sample it diffuses inside 3 Place the drop on the center of the slide IMPORTANT Always make a drop at the end of the FUJI clean tip and then apply the sample to the slid...

Page 93: ...button when the sample has already been filtered by the PF If the Start button is tapped the analyzer may malfunction Precautions for samples NOTE Use whole blood samples that have a hematocrit value Hct of 20 to 55 Up to about 350 μL of plasma including 85 μL of dead volume can be filtered The amount of filtered plasma decreases as the hematocrit value increases In the case of a hematocrit value ...

Page 94: ...F The PF card contains information to compensate for these differences 2 Using PF cards One PF card is included in one PF box with 50 filters Read a PF card in the same way a QC card is read Be sure to read the PF card when changing to a new box of PF The imported pro duction number lot is displayed on the Measurement preparation screen Make sure that this number matches the production number lot ...

Page 95: ...epare the consumables Prepare the tips mixing cups diluent and refer ence fluid as required Loading tips è Refer to Section 5 2 Loading mixing cups è Refer to Section 5 3 Loading diluent è Refer to Section 5 4 Loading reference fluid è Refer to Section 5 5 3 Load a PF sample rack on the sample disk and then load the slides 4 ...

Page 96: ...ng tube when the PF is loaded There must be empty space in the tube Refer to Section 4 4 2 2 Mix the whole blood sample in the blood collecting tube by turning it over five or six times NOTE Make sure that bubbles do not form 3 Remove the cap from the blood collecting tube and then put it on the PF sample rack 4 Insert a PF into the blood collecting tube Make sure the PF does not rise from the sam...

Page 97: ... PF and then remove the sample container WARNING When discarding used PF always follow biohazard procedures wear gloves a lab coat and safe ty goggles in accordance with the sample handling rules of your facility If any part of the body comes in contact with a sample immediately rinse the contaminated body part thoroughly under running water and then use ethyl alcohol as a disinfectant Seek medica...

Page 98: ...ample suction nozzle of the PF to be physically blocked by additives or other materials NOTE As shown in the illustration below the required sample volume is 6 5 mL or more for Ø16 100 mm tubes 6 mL or more for Ø13 100 mm tubes and 3 mL or more for Ø13 75 mm tubes 0023 PF 10mmの空き スペースが必要です 10mmの空き スペースが必要です 測定可能範囲 測定可能範囲 検体は φ16x100mm管で6 5ml以上 φ13x100mm管で6ml以上 必要です 検体は φ13x75mm管で3ml以上 必要です 100mm 7...

Page 99: ...eviously filtered by the PF IMPORTANT The retest button is designed only to automatically enter the sample number and sample ID of the previous sample position Do not use this button to retest samples from before the previous sample CAUTION Do not tap the START button when the sample has already been filtered by the PF If the START button is tapped the analyzer may malfunction Retesting after filt...

Page 100: ...hen load the consumables and sample è Refer to Section 3 3 2 Tap the retest button on the Meas urement results screen 3 Tap the sample rack display of the sam ple position to be retested NOTE The sample information registration screen displays the previous sample number or sample ID that was meas ured at the Sample A position NOTE Tap the Dilution button to set the dilu tion factor 4 Tap the OK bu...

Page 101: ...isplay 5 Close the sample cover and then lock it with the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can be opened 6 Tap the START button to start retesting NOTE When a test is rerun PF filtering process is not performed 4 ...

Page 102: ...suring CM and ISE slides 電解質 一般比色 図 4 5 2 1 ISE CM slide We recommend loading an ISE slide last If an ISE slide is measured first it might take extra time for measurement If a CM slide is loaded after an ISE slide an extra tip will be used 3 Measuring CRP CM and ISE slides CRP 電解質 一般比色 図 4 5 2 4 ISE CRP CM slide Stack CM slides before CRP and ISE slides 4 Measuring CM slides that require dilution ...

Page 103: ...f patients for the sam ples to be measured from DMS The Work list index screen appears NOTE If there are no measurement requests the buzzer sounds and the message There is no request information is dis played NOTE The display of the list is configured by Work list selection setting in Mode function for administrator è Refer to Section 9 3 8 NOTE On DMS Sample No is Sample ID and Sample ID is Patie...

Page 104: ...struction Manual 2022 04 2 Tap the sample rack for which you want to register sample information 3 Tap the sample information you want to register NOTE If the Select all button is tapped infor mation is registered in order from the first sample information 4 ...

Page 105: ...he work list measurement items can be printed by tapping the button NOTE If the Deselect all button is tapped registrations for all racks are cleared 4 Tap the OK button to finish registration The Measurement preparation screen appears NOTE With the software version 2 0 or later is installed a measurement request can be registered to multiple sample racks In this case is indicated on the right han...

Page 106: ...licable usage methods 897N201058J FDC NX700 Instruction Manual 2022 04 START STOP 0 3 A B D C E MYID001 0160503010 0160503009 NOTE Work lists can also be obtained from the sample information registration screen 4 ...

Page 107: ...sample types Sample types can be switched as batch all of sample rack A through E 1 Tap the Function menu button 2 Tap the PS W or U button for sam ple type switching NOTE This setting is applied to sample rack A through E 3 All sample type displays on the measure ment preparation screen are switched 4 ...

Page 108: ... START STOP 0 3 A B D C E MYID001 0160503010 0160503009 2 Check the sample type setting Make sure the sample type is Plasma Serum If it is not tap the Sample type button IMPORTANT The sample type of the sample slide and the sample type setting of the FDC NX700 must be same Otherwise an incorrect measurement result will be obtained 3 Tap the OK button to enter the meas urement conditions of Sample ...

Page 109: ...p sample rack display B START STOP 0 3 A B D C E MYID001 0160503010 0160503009 5 Tap the Sample type button and then choose Urine 6 Tap the OK button to confirm the meas urement conditions of Sample B The sample rack display lights up green 7 Tap the START button to start measure ment 4 ...

Page 110: ...number tap the No button at the measurement results screen Refer to Section 3 4 2 Enter the administrator password and then tap the OK button The Sample number entry screen appears 3 Edit the sample number and then tap the OK button The analyzer returns to the measurement results screen NOTE You can also edit the sample ID using the same procedure IMPORTANT Be sure not to make any entry errors NOT...

Page 111: ...may be read Make sure that the barcode data sample ID printed in the meas urement results are correct NOTE Carefully read the instructions included with the sample barcode reader before using it A maximum of 13 alphanumerical characters can be input for a sample ID Always use the sample barcode reader under lighting at 3 000 lux or lower and avoid direct sunlight NOTE Always connect the cable of t...

Page 112: ...he barcode of the sample using the barcode reader After reading is complete the analyzer makes a beep sound The sample ID is displayed on the LCD for the selected sample and the request information is registered NOTE Position the sample barcode reader properly facing to the barcode label Otherwise an incorrect reading may oc cur NOTE Make sure the displayed ID is correct NOTE When reading a barcod...

Page 113: ...D using the sample barcode reader After reading is complete the analyzer makes a beep sound The operator ID is displayed on the LCD and registered NOTE Position the sample barcode reader properly facing to the barcode label Otherwise an incorrect reading may oc cur NOTE Make sure the displayed ID is correct NOTE To re input a barcode tap the Operator ID button and then re read the barcode 4 ...

Page 114: ......

Page 115: ...splay of each consumable to display the loading screen When the consumable loading is complete press the Finish button to exit NOTE If there are not enough consumables tips mixing cups diluent reference fluid etc sam ple application may be interrupted or cancelled NOTE Load consumables while the analyzer is not applying samples 5 ...

Page 116: ... to interrupt application refer to Sec tion 4 2 and then load the tips It is unsafe to set tips during sample application due to the mov ing parts of the sampler unit and the sample disk Tips 1 With Warming up Lamp warming up or Ready to test displayed on the LCD make sure that the analyzer is not apply ing samples 2 Tap the tip display on the LCD 3 The Tip loading screen appears The sample disk t...

Page 117: ...e the tip racks from the sample disk before loading tips 5 Load tips in all holes of the tip racks NOTE Always set tips in all holes of the tip racks Tip rack loading positions 6 Load the tip racks on the sample disk 7 To initialize the tip usage position tap the Reset button 8 Tap the Finish button to exit The sample disk turns and the sample rack is moved to the front 5 ...

Page 118: ...hen the analyzer is applying samples tap the PAUSE button to interrupt sample application refer to Section 4 2 and then load the mixing cups It is unsafe to set tips during sample application due to the moving parts of the sampler unit and the sample disk WARNING Always wear gloves when handling used mixing cups to avoid biohazards Mixing cup display NOTE The quantity of mixing cups that can be us...

Page 119: ...ING Used mixing cups are infectious industrial waste Process this waste in compliance with any appli cable regulations in your country such as by incin eration melting sterilization or disinfection 5 For the mixing cups that were replaced tap the Reset button to reset the quan tity and then tap the Finish button to exit The sample disk turns and the sample rack moves to the front 5 ...

Page 120: ...unit and the sample disk IMPORTANT Use specified tube for the diluent Otherwise measurement errors or malfunctions may occur The specified tubes for diluent are the FUJI PLAIN TUBE 1 5 mL and the Ø13 75 mm test tube IMPORTANT The diluent for CRP measurements evaporates progressively and the concentration of the fluid may affect the measurement results Replace Ø13 75 mm test tubes within 12 hours R...

Page 121: ... the diluent display on the LCD 4 The Diluent loading screen appears The sample disk turns and the consumables are moved to the front CAUTION Do not touch the sample disk until it stops 5 Load the tube containing diluent in ac cordance with diluent label of the sample disk IMPORTANT Be sure to load diluent in accordance with the diluent label Otherwise the measurement results may be negatively aff...

Page 122: ...5 8 5 Loading consumables 897N201058J FDC NX700 Instruction Manual 2022 04 6 Tap the Finish button to exit The sample disk turns and the sample rack is moved to the front 5 ...

Page 123: ...ation refer to Section 4 2 and then load the reference fluid It is unsafe to set tips during sample application due to the moving parts of the sampler and the sample disk IMPORTANT Use a FUJI PLAIN TUBE 0 5 or 1 5 mL for reference fluid Otherwise measurement errors or malfunctions may occur Configure the type of container to be used via Tube setting for diluent reference fluid in Mode function for...

Page 124: ...lying samples 3 Tap the reference fluid display on the LCD 4 The Reference fluid loading screen ap pears The sample disk turns and the consumables are moved to the front CAUTION Do not touch the sample disk until it stops Reference fluid 5 Load the tube containing the reference fluid IMPORTANT Clean the cap of the reference fluid every time you load the reference fluid è Refer to Section 6 8 5 ...

Page 125: ...umables 897N201058J FDC NX700 Instruction Manual 2022 04 Sealing NOTE Make sure the reference fluid cap is closed properly 6 Tap the Finish button to exit The sample disk turns and the sample rack is moved to the front 5 ...

Page 126: ......

Page 127: ...intaining the analyzer get training from our technical support staff CAUTION Refer to each section for information about usable solvents for cleaning and decontamination CAUTION Do not use alcohol to clean the sampler cover translucent Otherwise the surface will be damaged IMPORTANT After cleaning always connect the power cord switch on the power and then check operation IMPORTANT If you do not pe...

Page 128: ...3 ISE unit Once every 3 months or as appropriate such as when inaccurate measurement values occur Refer to Section 6 3 Recording paper Appearance of red lines on both sides of paper Refer to Section 6 4 Light source lamp When a photometer gain setting error occurs or the lamp s cumu lative illumination time exceeds 1000 hours Refer to Section 6 5 Sampler O ring Once a month Once a year Refer to Se...

Page 129: ...f parts in the follow ing table Parts Replacement interval Interference filter Once every 4 years Syringe Once every 3 years Incubator pressure plate Inspection once every 3 years WARNING After inspection always connect the power cord switch on the power and then check operation NOTE This device is expected to perform about 16 000 standard usage tests a year 67 tests per day when used 240 days a y...

Page 130: ...PORTANT If the air filters are not cleaned an increase in the temperature inside the analyzer may negatively affect the measurement values NOTE Clean all 4 filters 1 Switch off the analyzer power Louver Louver 1 2 2 Remove the four louvers and pull out the air filters located inside of the louvers Lift the louver upward 1 and then pull forward 2 to remove 6 ...

Page 131: ... water NOTE If you wash the filter with running water make sure it is dried well before setting in the analyzer 1 2 4 Insert the filter into the louver and put the louver back in the analyzer Push up the louver 1 and then push it toward 2 IMPORTANT If the analyzer is used without setting the filters measurement values may be negatively affected 6 ...

Page 132: ...affected by stains from blood samples etc inside of the incubator and ISE unit These parts should be checked and cleaned at least once every three months When a slide transfer error or a tip ejecting error occurred clean the sample application unit 6 3 1 Reference plate level check Perform Reference plate level check before cleaning 1 Tap the button 2 Tap the Normal mode button 6 ...

Page 133: ...ppears 4 Check the results of Reference plate level check When has appeared in the check results cleaning is necessary Tap the OK button Using the procedure on the following pages clean the transfer bar incubator sample application unit and ISE unit In the case of Level 1 soiling 0 49 In the case of Level 2 soiling 50 79 In the case of Level 3 soiling 80 or higher 6 ...

Page 134: ...anual 2022 04 6 3 2 Cleaning the transfer bar incubator sample application unit and ISE unit 1 Cleaning the transfer bar 1 Tap the Cleaning mode button in Normal mode 2 Tap the Start Cleaning button NOTE Keep your fingers clear as the sample disk moves 6 ...

Page 135: ...ull the sample application unit cover frontward to remove it Slide transfer weight 6 Remove the two slide transfer weights NOTE There are 2 types of slide transfer weight one each for CM tests and ISE tests The slide transfer weight for CM tests has hooks on the left so lift up from the right before remov ing it When installing it insert the hooks before loading it For CM For ISE 6 ...

Page 136: ...e application unit 8 Next clean the incubator sample appli cation unit and ISE test unit NOTE After cleaning attach the removed slide transfer weight and sample application unit cover at their original positions and then tap the button to return to the previous screen CAUTION Keep your fingers clear as the sample disk moves 9 Tap the button to switch off the power CAUTION Keep your fingers clear a...

Page 137: ...ncubator Sampler Panel unit pedestal 1 Open the sampler cover and panel unit Move the sampler unit to the right and then remove the panel unit pedestal NOTE Be careful not to open the reference fluid cap Connector 2 Unplug the connector of the incubator cable low end by turning it to the left Incubator cover 3 Remove the incubator cover 6 ...

Page 138: ... plates while the incubator backside pressure plate surface is facing up Wipe the incubator cells in the same way IMPORTANT Do not use a cotton swab The components of the cotton swab may negatively affect the measure ment values IMPORTANT Do not touch the surface of the pressure plates with your bare hands Otherwise measurement values may be negatively affected IMPORTANT Moisten gauze or a similar...

Page 139: ...ence plate and clean the white plate and black plate with a dry cotton swab Photometer 8 Clean the photometer head with a dry cotton swab Thumbscrew Reference plate 9 Load the reference plate Tighten the 2 thumbscrews evenly IMPORTANT Tighten the two thumbscrews securely Otherwise the measurement results may be negatively af fected 6 ...

Page 140: ...ation unit 2 Wipe the sample application unit with gauze or a cotton swab moistened with lukewarm or cold water 4 ISE unit probe Probe Make sure there is no dust on the six ISE probes When you clean the ISE probes wipe off the dust using a dry cotton swab NOTE Do not use solvents such as methyl alcohol or isopropyl alcohol CAUTION ISE probes are needle shaped parts Do not touch them directly by ha...

Page 141: ... the slide transfer weights 2 Load the sample application unit cover and tighten the thumbscrews 3 Load the incubator cover 4 While aligned with the groove plug the connector of the incubator cable into the incubator and then lock it by turning it to the right 5 Load the panel unit pedestal IMPORTANT Tighten the thumbscrews securely Otherwise the measurement results may be negatively affected IMPO...

Page 142: ...ver 2 2 Remove the remaining recording paper 3 Load the new recording paper NOTE If the paper replacement was performed during a measurement the measure ment results might be printed out NOTE Set the paper roll so the paper is pulled from under the paper roll as shown below OK Incorrect NOTE If the paper is loaded in reverse no text is printed on the thermal paper 4 Pull the top end of the recordi...

Page 143: ...ar 1 000 hours CAUTION The light source lamp gets very hot Before cleaning the light source unit or replacing the light source lamp turn off the power wait at least 5 minutes and then check the lamp has cooled Front bottom cover 1 Pull out the disposal box and then re move the front bottom cover 2 1 3 Lamp unit Thumb screw Connector 2 Loosen the thumbscrew as shown in the left figure and remove th...

Page 144: ...ecure so that metal hooks hold from above the lamp housing 5 Turn the thumbscrew of the lamp unit to the left remove the lamp and then re place it with a new one IMPORTANT Tighten the thumbscrews securely Otherwise the measurement results may be negatively affected NOTE When you replace the lamp hold it by its base Do not touch the glass surface with bare hands 6 ...

Page 145: ...IMPORTANT Tighten the thumbscrew securely Otherwise the measurement results may be negatively affected 8 Attach the front bottom cover load the disposal box and then switch on the power of the FDC NX700 IMPORTANT Be sure to load the disposal box If the analyzer is used without the box the measurement results may be negatively affected 9 After starting the analyzer go to Lamp management in Normal m...

Page 146: ...ear gloves a lab coat and safety goggles If you accidentally touch a contaminant immediately rinse the contaminated area thoroughly un der running water and then disinfect the area Seek medical assistance if necessary IMPORTANT If you do not inspect and replace the sampler O ring the sample application volume will be inac curate which may negatively affect the measurement values 6 6 1 Inspecting t...

Page 147: ...pler leak check 4 Follow the on screen instructions to pre pare for the leak check 5 Tap the nozzle to be checked to select it and then run the leak check Once check ends the Check results screen ap pears CAUTION Keep your fingers away as the sampler unit and sample disk move 6 ...

Page 148: ...O ring NOTE The check results can be printed by tapping the button If the check results are NG tap the Replace O ring button and then replace the O ring according to the following procedure NOTE The check results can be printed by tapping the button 7 Once the check ends open the sampler cover and then remove the sampler leak check tool mounted on the sampler noz zle by hand 6 ...

Page 149: ...ORTANT If you do not replace the O ring periodically the sample application volume will be inaccurate which may negatively affect the measurement values IMPORTANT Be sure to use the O ring designed for the FDC NX700 Otherwise the sample application volume will be inaccurate and the measurement results may be negatively affected 1 Remove the sample rack and slide car tridge 2 Tap the button 3 Tap t...

Page 150: ...he nozzle on which you want to replace the O ring to select it The sampler unit moves and the sampler noz zle comes down CAUTION Keep your fingers away as the sampler unit moves 6 Follow the on screen instructions to replace the O ring If this button is tapped the following expla nation appears 6 ...

Page 151: ...6 25 6 Maintenance 897N201058J FDC NX700 Instruction Manual 2022 04 7 Tap Leak check to check for leaks è Refer to Section 6 6 1 6 ...

Page 152: ...oiling may be the cause Clean the slide reading section to avoid these errors 1 Switch off the analyzer power 2 Open the sampler cover and panel unit 3 Remove the slide cartridge of sample A 4 Turn the sample disk by hand so sample A is in the aspiration position 5 Use an air blower through the opening for placing the slide cartridge of sample A as shown at left 6 ...

Page 153: ...erence fluid will be concen trated which may negatively affect the measurement values 1 If reference fluid is loaded remove the reference fluid tube and discard it Reference fluid cap packing 2 Remove the reference fluid cap packing as shown in the illustration on the left Cotton swab Reference fluid cap packing 3 Wipe off the reference fluid cap packing using a cotton swab or gauze moistened with...

Page 154: ...e 897N201058J FDC NX700 Instruction Manual 2022 04 Reference fluid cap packing 4 Set the reference fluid cap packing prop erly NOTE Do not touch the bottom side reference fluid side of the reference fluid cap packing 6 ...

Page 155: ... 2022 04 6 9 PF unit maintenance 6 9 1 Cleaning the PF unit sensor reference plate and suction pad Clean when errors occurred during PF filtering 1 Remove the sample rack and slide car tridge 2 Tap the button 3 Tap the Normal mode button 4 Tap PF maintenance 6 ...

Page 156: ...the OK button CAUTION Keep your fingers away as the sample disk and PF section move 7 Clean according to the on screen in structions Surface of PF sensor glass surface PF Suction Pad black 8 Return the tip racks to their original po sition and then tap the OK button to end cleaning Sample disk turns 6 ...

Page 157: ... 897N201058J FDC NX700 Instruction Manual 2022 04 6 9 2 Inspecting the PF Suction Pad The PF Suction Pad must be inspected monthly and replaced yearly 1 Tap the button 2 Tap the Normal mode button 3 Tap PF maintenance 6 ...

Page 158: ...check tool on Sample rack A and then tap the START button to run the leak check If check ends the Check results screen appears CAUTION Keep your fingers away as the sample disk and PF section move 6 Check the results of the Leak check If the check results are OK it is not necessary to clean or replace the pad NOTE The check results can be printed by tap ping the button 6 ...

Page 159: ...If this does not resolve the problem tap the Procedure button and then check the attach ment or replace the PF pump tube to using the procedure on the following page NOTE If the Result details button is tapped the Advanced leak check results display screen appears NOTE The check results can be printed by tap ping the button 7 Once the check ends remove the PF sample rack and PF leak check tool 6 ...

Page 160: ...tious waste in compliance with any applicable regulations in your country Tips 1 With Warming up Lamp warming up or Ready to test displayed on the LCD make sure that the analyzer is not apply ing samples 2 Tap the tip display on the LCD 3 The Tip loading screen appears The sample disk turns and the consumables are moved to the front CAUTION Do not touch the sample disk until it stops 6 ...

Page 161: ...Instruction Manual 2022 04 Tip rack 4 After the sample disk stops remove the tip racks from the sample disk 5 Tap the Finish button to exit The sample disk turns and the sample rack is moved to the front Power button 6 Press the power button 6 ...

Page 162: ...eft 9 Turn the PF Suction Pad in the direction shown and remove it while loosening 10 Install the new PF Suction Pad by turn ing it in the opposite direction NOTE Be sure to tighten the PF Suction Pad securely 11 Start the analyzer 12 Run the leak check If the check results are NG perform the re placement again è Refer to Section 6 9 2 6 ...

Page 163: ...analyzer and unplug the power cord Replace the PF Tube once a year Replacement is also necessary if the PF has not been used for one month or longer 1 Remove the disposal box and remove the front bottom cover Cover B Screw A 2 Loosen the screw A and remove the cover B Holder C Tube D 3 Slide the holder C to the left and re move the tube D from the rotor Front bottom cover 6 ...

Page 164: ...r NOTE As shown in the diagram to the left align the base of the joint E with the cover hole F NOTE Set so that the two front marks face the front Front Marks Cover hole F Joint E Joint E Cover hole F When attaching the cover B make sure that the hooks are hooked and interlocked NOTE If attachment is incorrect a W0134 PF sampling error may occur 6 Reattach the cover B using the screw A Joint G Red...

Page 165: ...58J FDC NX700 Instruction Manual 2022 04 8 Attach the front bottom cover 9 Attach the disposal box 10 After the replacement run the leak check If the check results are NG perform the re placement again è Refer to Section 6 9 2 6 ...

Page 166: ......

Page 167: ...slide and then meas ure 7 3 Monitoring results Configure the target values and acceptable ranges for measurements from control fluid results and monitor for changes over time 7 4 Control mode You can measure control liquid by using Control of the Function menu to configure the entered cor relation coefficients a and b to a 1 and b 0 temporarily This state is called Control mode è Refer to Section ...

Page 168: ...ontrol 7 2 897N201058J FDC NX700 Instruction Manual 2022 04 7 6 Control chart 1 Tap the button 2 Tap the Control button 3 The measurement preparation screen of control mode appears 4 Tap the graph display button 7 ...

Page 169: ... changed Control lot was changed Outside of accept able range 6 The control chart appears NOTE If the control liquid button is tapped you can change the control liquid to be dis played NOTE Measurement results of outside of measurement range disposal box open and expired slide are not reflected in the control chart NOTE If the measurement date is touched and selected and the show hide button is to...

Page 170: ...n outside of the acceptable range of control Acceptable range lower limit Acceptable range upper limit NOTE The acceptable range of control displays the value entered for Range setting for control Refer to Section 9 3 17 NOTE In case of ALPi and LDHi slides the display of Test name will not follow the Test item name setting but ALPi PS or LDHi PS But the test name of each measure ment result will ...

Page 171: ... or warnings appear in the measurement results the measurement results may be inaccurate Refer to the related troubleshoot ing pages and rerun the measurement 8 1 Error display If an error occurs an error message appears on the LCD Example of display Error code Content of error message Additional information NOTE Some error codes may show additional infor mation Example 1050 GLU P slide code 12000...

Page 172: ...uction Manual 2022 04 8 1 1 Error log You can view the history of errors that occurred by using Error log in the function menu 1 Tap the button 2 Tap the Error log button An error log list is displayed 3 Tap an error The error details are displayed 8 ...

Page 173: ...8 Troubleshooting 8 3 897N201058J FDC NX700 Instruction Manual 2022 04 8 ...

Page 174: ...154 Mixing cup error 8 2 4 3 8 13 E0200 E0204 E0210 E0213 E0220 E0227 Slide transfer error 8 2 6 8 22 E0300 E0301 E0302 Reference white plate measurement error during initial operation 8 2 8 8 26 E0509 E0510 Signal system error inside circuit board 8 2 8 8 26 E0511 Light intensity error 8 2 5 1 8 20 E0515 E0516 E0517 Signal system error inside circuit board 8 2 8 8 26 E0530 E0531 ISE measurement u...

Page 175: ... E4200 E4201 Calibration error 8 2 14 3 4 8 41 E4202 E4203 Calibration error 8 2 14 5 8 42 E4204 Calibration error 8 2 14 6 8 42 E4205 Calibration error 8 2 14 5 8 42 E4401 E4403 Signal system error inside circuit board 8 2 8 8 27 E4410 Measurement result reference error 8 2 8 8 27 E4411 Control chart reference error 8 2 8 8 27 E4454 E4475 E4478 E4481 E4490 E4600 E4613 Signal system error inside c...

Page 176: ...error 8 2 4 7 8 16 W0146 Liquid volume error 8 2 4 9 8 18 W0151 W0152 Mixing cup error 8 2 4 3 8 13 W0170 QC card has not been read 8 2 9 2 8 31 W0173 The slide information has not been read 8 2 3 8 11 W0180 Calibration has not been performed with an im muno chemical test 8 2 14 1 8 40 W0190 W0191 Calibration error 8 2 14 2 8 40 W0214 Slide transfer error 8 2 6 8 22 W0500 W0503 Malfunction of the ...

Page 177: ...511 Type 4 LAN communication error 8 2 13 1 8 39 W4520 Error related to work list 8 2 13 1 8 39 W4521 Communication warning 8 2 13 1 8 39 W4720 Clock time error 8 2 8 8 29 W4721 W4731 W4733 W4735 W4736 Signal system error inside circuit board 8 2 8 8 30 W5000 W5020 Signal system error inside circuit board 8 2 8 8 30 W5018 W5019 USB memory read error 8 2 8 8 30 W5050 USB memory has not been inserte...

Page 178: ...the up per limit of the measurement range Refer to Section 8 2 11 OR Measurement value of the ISE test is lower than the lower limit of the measurement range Refer to Section 8 2 11 Temperature control error during measurement Refer to Section 8 2 7 Light intensity is fluctuating during measurement Refer to Section 8 2 5 E Unable to calculate the measurement result because measurement was interrup...

Page 179: ...nstalled analyzer will show for the ISE slides that expire 2023 or later or of lot number 130000 to 139999 NOTE Depending on the device settings there may be cases in which is not displayed and an expired slide cannot be measured Problem with sample application on slide not applied Rerun the measurement Supplement Measurement range The measurement range determination range and the reference interv...

Page 180: ...ase contact your dealer for assistance 2 White lines appear on printed characters or certain parts of characters are missing The printer head has malfunctioned Please contact your dealer for assistance 3 Printer errors Error code Error message Error condition Troubleshooting W5110 Printer error Arm open No paper Set paper No recording paper Load recording paper onto the printer è Refer to Section ...

Page 181: ...ip can be detected on the sample nozzle after mounting The tip is not loaded correctly Load a tip correctly and then rerun the test from the beginning è Refer to Section 5 2 NOTE Set tips into all holes of the tip racks after this error has oc curred E0083 Reference fluid tip detection ERR No tip detected Set tips and rerun the tests from the begin ning No tip has been de tected on the refer ence ...

Page 182: ...rding the tip at the last startup Please check the dis posal box If the disposal box is not empty please shut down empty the dis posal box and restart Please check the disposal box If the disposal box is not empty please shut down empty the disposal box and restart W5402 Reconfirmation An error occurred while discarding the tip at the last startup Please check that the sample application unit cove...

Page 183: ...rror still occurs after switching the analyzer off and on please contact your dealer for assistance 4 During a sample volume error Error code Error message Error condition Troubleshooting W0120 Sample shortage Sample is insufficient Check the sample Either the sample is not loaded or the sam ple volume is insuffi cient Make sure that the sample volume in the tube is within the measurable range and...

Page 184: ...t E0112 Spotting clogging ERR sample Clogging was detected during sampling Check the sample A clog was detected during sample applica tion ejection E0125 Spotting clogging ERR diluted sample Clogging was detected during sampling Check the sample and diluent A clog was detected during sample applica tion ejection The sample may contain fibrin deposits After checking the sample rerun from the measur...

Page 185: ...RI CHEM AUTO TIPS are being used b Make sure that tips are not being reused c Make sure that there are no bub bles on the sample surface in the sample tube d Make sure that a sample rack that matches the blood collecting tube or FUJI tube is being used e Inspect the sampler O ring Refer to Section 6 6 1 IMPORTANT Measurement values with an mark might be incorrect Run the measure ment again IMPORTA...

Page 186: ...to Section 5 5 NOTE Replace the diluent reference fluid container with a new one Do not add diluent or reference fluid W0142 Surface detection ERR reference fluid Reference fluid volume is too much Check the reference fluid volume There is too much ref erence fluid volume W0123 Diluent is insufficient Diluent is insufficient Check the diluent There is not enough diluent W0143 Diluent is insufficie...

Page 187: ...he diluent and reference fluid and then rerun from the measurement request NOTE If the error occurs again there might be a piping system prob lem Please contact your dealer for assistance E0122 Spotting clogging ERR diluent Clogging was detected during sampling Check the diluent Clogging was detected during diluent applica tion E0127 Aspiration clogging ERR diluent Clogging was detected during sam...

Page 188: ...ibrator Check the calibrator liquid volume and then measure again W0145 Insufficient calibrator Calibrator is insuffi cient Check the calibrator volume and perform calibration or verify calibration again W0146 Surface detection ERR CP Calibrator volume is too much Check the calibrator volume and perform calibration or verify calibration again There is too much cali brator Make sure that the cap of...

Page 189: ...1002 Sampler up down motor operation ERR Problem on the sample nozzle side Perform inspection Contact your dealer E1010 E1011 E1013 W1014 Sampler up down motor ERR Problem on the refer ence fluid nozzle side Perform inspection Contact your dealer Movement error of reference fluid nozzle E1012 Sampler up down motor operation ERR Problem on the refer ence fluid nozzle side Perform inspection Contact...

Page 190: ...Clean the reference plate è Refer to Section 6 3 Clean the photometer è Refer to Section 6 3 After cleaning switch on the power If the errors still occur after cleaning replace the light source lamp è Refer to Section 6 5 NOTE If the error occurs again please contact your dealer for assis tance W0030 Lamp replacement WARNING Low light intensity Clean photometer The light intensity of the light sou...

Page 191: ... 5 minutes and then check the lamp has cooled 3 During an interference filter motor error Error code Error message Error condition Troubleshooting E0900 E0901 E0903 W0904 Filter motor ERR Problem with the inter ference filter Contact your dealer Movement error of the filter motor Switch the analyzer off and on NOTE When the error is not resolved please contact your dealer for assistance E0902 Filt...

Page 192: ...gain please contact your dealer for assis tance E0210 E0213 W0214 Eject motor ERR Problem with the slide transfer system Perform inspection Contact your dealer W1652 Initialization ERR Problem with the slide transfer system Perform inspection Contact your dealer E0600 E0604 Incubator motor ERR Problem with the incu bator Perform inspection Contact your dealer Movement error of the incubator motor ...

Page 193: ...ment error of the disk motor of the sam ple disk section The sample application unit or the incu bator may be dirty or the disposal box may be full of slides and tips Switch off the analyzer and perform the following Cleaning the disposal box Cleaning the incubator è Refer to Section 6 3 Cleaning the sample application unit è Refer to Section 6 3 NOTE If the error occurs again please contact your ...

Page 194: ...mperature b Switch off the power c Clean the air filter è Refer to Section 6 2 d Make sure that thumbscrews of the incubator are tightened properly and the incubator cable is connect ed to the connector properly èRefer to Section 6 3 3 e Make sure that the sample ap plication unit cover is attached to the analyzer properly Tighten the thumbscrews securely f Leave the analyzer for about 10 minutes ...

Page 195: ... again NOTE After inserting the connector of the incubator cable you must turn it to the right to lock it è Refer to Section 6 3 3 NOTE If the error occurs again please contact your dealer for assis tance E1503 Temperature control ERR Incubator temperature sensor damaged Contact your dealer E1504 Temperature control ERR Incubator heater dis connected Check the incubator cable E1512 ISE temperature...

Page 196: ...Faulty A D controller Contact your dealer E0515 DA conversion ini tialization ERR Contact your dealer E0516 AD conversion reset ERR Contact your dealer E0517 AD conversion ini tialization ERR Contact your dealer E0532 E0533 E0537 E0538 ISE self test AD con version ERR For ISE measurement turn the power SW OFF and ON Contact your dealer in case the ERR repeats Other parameters besides ISE can be me...

Page 197: ...ase contact your dealer for assistance E1750 E1755 Slide reading unit failure Contact your dealer If the error still occurs after switching the analyzer off and on please contact your dealer for assistance E3900 Internal communica tion ERR Contact your dealer E4401 E4403 Save ERR Contact your dealer E4410 Measurement result reference error The number of meas urement results with the new result cur...

Page 198: ...tting for reference intervals is correct W4465 W4466 W4468 Memory recovery WARNING Recovery performed from backup memory Check whether the QC card is read by con firming the lot number If not re read the QC card Check the lot information whether the QC card has been read If the card has not been read read the QC card again W4467 Memory recovery WARNING Recovery performed from backup memory Check w...

Page 199: ...o the default Contact your dealer in case the ERR repeats If the error occurs again please contact your dealer for assistance E4490 File deletion ERR Contact your dealer if this recurs Please contact your dealer for assistance E4600 E4613 Calculation ERR Contact your dealer W4720 Clock setting WARN ING Set date time If the error still occurs contact your dealer The date and time have been reset Se...

Page 200: ...stance W4736 LCD brightness set ting WARNING Contact your dealer W5000 USB memory read ERR Contact your dealer W5018 USB memory read ERR Operator ID set ting Lack of necessary files in the USB memory Check the USB mem ory and then reselect a menu Contact your dealer in case the ERR repeats Lack of necessary files in the USB memory Check the USB memory and retry ac cording to the error message If t...

Page 201: ...ftware error has occurred Please contact your dealer for assistance 8 2 9 QC card PF card and DI card errors 1 When data is not displayed after a QC card is read Error code Error message Error condition Troubleshooting Problem reading QC card information Read the QC card again after making sure its correct orientation If it does not improve insert the air blow nozzle into the QR code reader slot a...

Page 202: ...ded with the PF 4 When a DI card has not been read Error code Error message Error condition Troubleshooting W4012 No DI info No DI info found Can not read this card Contact your dealer There is no DI card information Please contact your dealer for assis tance 5 Other precautions Error code Error message Error condition Troubleshooting W1700 W1702 W1704 W1706 QC card reader failure QC card reader d...

Page 203: ...rs Error code Error message Error condition Troubleshooting W0040 Incorrect sample WARNING The sample type set in analyzer does not correspond to the read sample type Set the sample type by SAMPLE key and the slide s sample type corresponding to the sample Rerun the tests from the begin ning The sample type se lected by the Sample type button and the sample type of the slide do not match Example W...

Page 204: ...irection ERR Incorrect loading direc tion of ISE slide Check the ISE slide The ISE slide is loaded in reverse Load the ISE slide in the slide cartridge in the proper direction and then meas ure again W1654 Check sample rack Unusable sample rack was set in place Check the sample Contact your dealer in case the ERR repeats A sample rack that cannot be used is loaded Use a sample rack intended for th...

Page 205: ...gain ERR For ISE measurement turn the power SW OFF and ON Contact your dealer in case the ERR repeats Other parameters besides ISE can be measured An error was detected during a self test per formed before the start of an ISE measure ment The analyzer must be restarted for an ISE measurement If the error occurs again please contact your dealer for assistance Tests other than ISE do not require a r...

Page 206: ...lfunction Drift error and impedance error Impedance error and outside of the measurement range Drift error and outside of the measurement range Drift error impedance error and outside of the measurement range Impossible measurement When the error code is not E3 The sample or reference fluid may have not been sufficiently absorbed into the slide Perform the fol lowing troubleshooting steps 1 Inspec...

Page 207: ...ion 8 2 12 PF errors Error code Error message Error condition Troubleshooting E0132 PF pressure ERR Replace the PF pump tube A pressure error oc curred during the filter ing process Replace the PF Tube è Refer to Section 6 9 4 E0133 E0150 PF sensor ERR Contact your dealer A PF sensor error has occurred Please contact your dealer for assis tance W0134 PF sampling ERR Error occurred during PF sampli...

Page 208: ...er Sensor error Contact your dealer A PF sensor error has occurred Please contact your dealer for assis tance E1302 PF motor operation ERR Problem with the plas ma filter Contact your dealer PF motor error move ment error Turn the power on and off If the error occurs again please contact your dealer for assistance W1651 Incorrect sample rack PF remeasurement requires a PF sample rack Check the sam...

Page 209: ...e the ERR repeats Type 4 LAN communi cation ERR Check the host computer and type 4 settings è Refer to Section 9 3 2 NOTE If the error occurs again please contact your dealer for assis tance W4520 Incorrect work list in formation Check whether work list information for the host computer is correct Contact your dealer in case the ERR repeats The received work list information does not match specifi...

Page 210: ...ue and re enter Barcode data error Check the input value and re enter 8 2 14 Calibration errors 1 An immuno chemical test has been performed without calibration Error code Error message Error condition Troubleshooting W0180 Calibration not per formed Perform calibration An immuno chemical test has been per formed without calibra tion Perform the calibration è Refer to Section 4 1 2 A slide not sui...

Page 211: ...id number of slides The number of meas ured slides for calibra tion data is insufficient or sample is unspotted This error may occur due to a lack of calibra tion data or no applica tion on the slide Rerun 2 measurements for the printed level W4220 Insufficient number of slides for verifying calibration Insufficient number of slides for verifying calibration Set the slides and re start verifying c...

Page 212: ...tion curve set ting ERR range error Misplacement of cali brator Check the level and expiration date of the calibrator and perform the calibration from the beginning E4205 Calibration calcula tion ERR Failed with error in calibration operation Do the calibration again Calibration calculation problem Calibrate again from the beginning 6 Measurement errors during calibration Error code Error message ...

Page 213: ... in progress The disposal box is full of slides and tips While a measurement is not being per form dispose the slides and tips in the disposal box NOTE When there are 150 slides or 75 tips in the disposal box this error occurs If the disposal box is pulled out it is determined that the slides and tips were discarded and the disposal count is reset Always empty the disposal box after pulling it out...

Page 214: ...5920 Turn the power ON 30 days or more have passed from when the analyzer was started For stable system performance turn the power ON A long time has passed since start up Switch the analyzer off and on W66xx Input value ERR Check input value and re enter The entered value is incorrect Check the entered value and enter it again 8 ...

Page 215: ...GPT ALT P 32 50 TCHO P 14 50 CPK P 33 50 NH3 P 15 50 LDH P 34 50 TG P 16 50 ALP P 35 50 CRE P 17 50 LAP P 37 50 TP P 18 50 CKMB P 38 50 ALB P 20 50 CHE P 39 50 TBIL P 21 50 AMYL P 43 50 Ca P 23 50 LIP P 44 50 IP P 24 50 TCO2 P 45 50 DBIL P 25 50 CRP S 51 50 HDLC P 26 50 Na 91 00 Mg P 28 50 K 92 00 Cl 93 00 GLU W 10 00 NH3 W 15 00 NOTE Slide codes are subject to change without notice Check the slid...

Page 216: ......

Page 217: ...l 9 1 START Manual pipetting 4 3 1 Normal mode Date Time settings 9 2 1 Brightness Volume Display brightness 9 2 2 Error sound Other general sound Lamp management Lamp ON time 9 2 3 Lamp off timer setting Lot information 9 2 4 Sampler leak check Sample nozzle 6 6 1 Reference fluid nozzle Replace O ring Sample nozzle 6 6 2 Reference fluid nozzle PF maintenance Cleaning 6 9 1 Leak check 6 9 2 Refere...

Page 218: ...selection setting Sample ID Patient name 9 3 8 Patient ID Patient name Correlation coefficients 9 3 9 Lot compensation coefficients 9 3 10 Standard curve coefficients 9 3 11 Lot qualification 9 3 12 PF calibration coefficients 9 3 13 Dilution setting 9 3 14 Tube setting for diluent refer ence fluid Diluent 9 3 15 Reference fluid Print setting for measured result Number of printing sheets 9 3 16 Pr...

Page 219: ...isabled a 1 b 0 of the function menu Measurement units follow Unit settings of the function menu Dilution conditions dilution is not normally performed This can be configured at the Sample information registration screen NOTE When the analyzer is switched on again control mode is canceled 1 Prepare the slide to be measured and then load it in the sample position to be measured 2 Place the control ...

Page 220: ...ect the control liquid from the list when nec essary 4 Select the control liquid lot from the list when necessary IMPORTANT When you configure the control liquid lot make sure it matches the lot of the control liquid to be used If it does not match the measurement may not be valid 5 For the sample type generally in the case of QP L QP H and QE select Plasma Serum 6 For dilution generally select No...

Page 221: ...mple cover and then lock it with the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can be opened 7 Tap the START button to start control measurement 9 ...

Page 222: ...sing control mode always tap the HOME button to clear control mode Display of measurement results Green when within the acceptable range of control Red when outside of the acceptable range of control Acceptable lower limit Acceptable upper limit NOTE The acceptable range of control displays the value entered for Control fluid setting see Section 9 3 17 9 ...

Page 223: ... Reference section Date Time settings 9 2 1 Brightness Volume Display brightness 9 2 2 Error sound Other general sound Lamp management Lamp ON time 9 2 3 Lamp off timer setting Lot information 9 2 4 Sampler leak check Sample nozzle 6 6 1 Reference fluid nozzle Replace O ring Sample nozzle 6 6 2 Reference fluid nozzle PF maintenance Cleaning 6 9 1 Leak check 6 9 2 Reference plate level check 6 3 1 ...

Page 224: ... power off will appear when is tapped to exit Normal mode After the shutdown reboot the analyzer 9 2 2 Brightness Volume You can configure the brightness of the LCD and buzzer sounds error warning measurement com pleted Configure the brightness of the LCD using Display brightness Use and to set the minimum and maximum brightness Configure the buzzer sound when an error oc curs by using Error sound...

Page 225: ...t the cumulative illumination time of the lamp after replacing the lamp If the Lamp off timer setting checkbox is select ed the lamp is switched off automatically after some time You can configure the time until the lamp is switched off NOTE The default setting is 20 minutes 9 2 4 Lot information The screen displays the lot information of currently registered QC cards You can also print all cur re...

Page 226: ...onnection LAN connection Calculated Parameter 9 3 3 Unit conversion 9 3 4 Switch measurement range display 9 3 5 Reference interval range setting 9 3 6 Sample No ID settings 9 3 7 Work list selection setting Sample ID Patient name 9 3 8 Patient ID Patient name Correlation coefficients 9 3 9 Lot compensation coefficients 9 3 10 Standard curve coefficients 9 3 11 Lot qualification 9 3 12 PF calibrat...

Page 227: ...r the setting to take effect The confirmation di alog for the power off will appear when is tapped to exit Mode function for adminis trator After the shutdown reboot the ana lyzer 9 3 2 Host connection setting Configure the connection terminal and communication parameters to connect to the host computer PC Furthermore switch the built in printer on or off When you will not connect to a host comput...

Page 228: ...utput from the printer You can choose from 4 communication types Type 1 Bidirectional connection Type 2 Unidirectional connection bidirectional format Type 3 Unidirectional connection standard format Type 4 Unidirectional connection HL7 for matted When was tapped the LAN configuration screen appears If an entry field is tapped the software keyboard refer to Section 2 3 appears so you can edit the ...

Page 229: ...ion Tap MLLP setting to set MLLP Minimal Lower Layer Protocol in communication Full MLLP is applied to full message in com munication Custom MLLP is applied to only at the end of message in communication Tap Sample type to select the sample type used in communication NOTE Rebooting the device is necessary for the setting to take effect The confirmation di alog for the power off will appear when is...

Page 230: ...T GOT value GPT value AST ALT AST value ALT value NOTE The calculation test name can be changed by Test item name setting in the Mode function for administrator è Refer to Section 9 3 20 Sodium Potassium ratio Na K Na K Na value K value Anion Gap Anion Gap mEq L or mmol L Anion Gap Na value Cl value TCO2 value IMPORTANT Calculation items display a value calculated from simultaneously measured valu...

Page 231: ...he same slide cartridge to perform the measurements If the Calculated Parameter button is tapped in Mode function for administrator precautions for use of calculation items are displayed To use calcula tion items after confirming the precautions tap the OK button Configure the calculation items to be used NOTE If the QC card of an item to be used for cal culation has not been read the calculation ...

Page 232: ... μkat L 0 0167 NH3 15 μg dL μmol L 0 7139 LDH 34 U L μkat L 0 0167 TG 16 mg dL mmol L 0 01129 ALP 35 U L μkat L 0 0167 CRE 17 mg dL μmol L 88 4 LAP 37 U L μkat L 0 0167 TP 18 g dL g L 10 CKMB 38 U L μkat L 0 0167 ALB 20 g dL g L 10 CHE 39 U L μkat L 0 0167 TBIL 21 mg dL μmol L 17 1 AMYL 43 U L μkat L 0 0167 Ca 23 mg dL mmol L 0 25 LIP 44 U L μkat L 0 0167 IP 24 mg dL mmol L 0 3228 TCO2 45 mmol L m...

Page 233: ...isplay method for data outside the measurement range passing the upper lower limit of the determination range for all tests If is configured an symbol is added to indicate the displayed measurement value is outside the measurement range If is configured a value outside the meas urement range is displayed by a or sym bol 9 ...

Page 234: ... 3 4 Furthermore you can set a reference inter val for each sample type P S W or U of a reference interval name Adding and editing of reference interval names Next page buttons Tap the button of the item you want to edit or add If an entry field is tapped the software keyboard refer to Section 2 3 appears so you can edit the test name and other details To change the sample type when configuring th...

Page 235: ...tion Manual 2022 04 Changing the display order Next page buttons Tap the button of the test for which you want to change the display order Tap the field of the test for which you want to con figure the transfer destination and then tap the OK button 9 ...

Page 236: ...onfigure whether to display the sample number in the measurement results under Edit and print No Configure whether to record the sequence num ber of the analyzer under Store No 9 3 8 Work list selection setting Configure whether to display the sample ID or pa tient ID at the Work list index screen 9 ...

Page 237: ... document IMPORTANT Incorrect entry of a and b causes incorrect measurement results Always make sure that the entries a b are correct Next page buttons Tap the Data input button NOTE If the button is tapped all correlation coefficient settings are printed NOTE If the Reset all tests button is tapped a reset confirmation screen appears To reset the correlation coefficients of all measure ment items...

Page 238: ...er the value To change the sample type tap the P S W or U button NOTE If the Reset button is tapped a reset confirmation screen appears To reset the selected measurement items and the cor relation coefficients of the sample type tap the OK button To cancel the reset tap the Cancel button NOTE If values are entered for a and b when values have already been entered the current values are not discard...

Page 239: ...nalyzer memory the FDC NX700 performs compensation calculations internally us ing the formula X Y b a In this way measured data obtained using the FDC NX700 are adjusted to match measured data that would be obtained using your instruments Y mg dl X mg dl FDC reference Y X Y 0 98X 5 0 10 600 0 600 A B Measurement using your instruments Method In order to obtain a better correlation it is es sential...

Page 240: ...above the meas urement range If the regression formula is Y 0 98X 5 A 5 mg dl B 607 mg dl Y mg dl X mg dl FDC reference Y 1 1X 5 Y X 0 10 600 0 600 A B Measurement using your instruments If the regression formula is Y 1 1X 5 A 14 mg dl B 550 mg dl Printout example Value under the meas urement range Value above the meas urement range If the measured data is below the measurement range an mark is ad...

Page 241: ...ber was changed even if the new values c d and e are entered by us ing this setting the new standard curve information will not be updated It is necessary to read the QC card when the slide type number was changed The slide type number is the most significant digit of a lot number Example The slide type number of Lot No 123456 is 1 Tap the button of the Measurement item sample type you want to edi...

Page 242: ...board refer to Section 2 3 appears so you can enter the value NOTE The most significant digit of a Lot number the type number is fixed Enter from the second digit NOTE Incorrect entries will cause incorrect meas urement results Always make sure that the displayed values are correct NOTE If the button is tapped all lot compen sation coefficient settings are printed NOTE For the expiration date for ...

Page 243: ...t results Always make sure that the entered coefficients are correct NOTE The calibration curve coefficients of this setting are entered automatically by QC card read ing and calibration Configuring the lot number 1 Tap the button of the measurement item for which the calibration curve is to be edited NOTE Only CRP items for which QC cards have been read are displayed List display area 2 From the ...

Page 244: ...e button is tapped the calibra tion curve coefficients are printed 4 First tap the Input field of Lot The Lot number entry screen software key board screen appears 5 Enter a lot number and then tap the OK button NOTE The most significant digit the type number is fixed Enter from the second digit The Calibration curve coefficient A entry screen software keyboard screen appears 9 ...

Page 245: ...rd screen appears 2 Enter B through F using the same proce dure Configuring the expiration date for use For the expiration date for use EXP enter xxxx xx four digits for the year and two digits for the month NOTE Also enter the hyphen IMPORTANT Incorrect entries will cause incorrect measurement results Always make sure that the displayed values are correct NOTE When the button is tapped values of ...

Page 246: ...abled It is possible to dis qualify slide LOTs which are registered in analyzer 1 Select a slide name which needs to change the setting 2 Dis qualification for selected slide LOT can be changed by selecting ON or OFF button NOTE When the disqualified slide lot is meas ured the test result will be indicated as CHK QC on screen print and in com munication 9 ...

Page 247: ...tapped all PF calculation correction coefficients are printed NOTE If the Reset all tests button is tapped a reset confirmation screen appears To reset the PF calculation correction coefficients of all measurement items tap the OK button To cancel the reset tap the Cancel button Tap the button of the measurement item you want to edit NOTE Only measurement items for which QC cards have been read ar...

Page 248: ...are keyboard refer to Section 2 3 appears so you can enter the value NOTE If the Reset button is tapped a reset con firmation screen appears To reset the PF calculation correction coefficients of select ed items tap the OK button To cancel the reset tap the Cancel button NOTE To cancel the configuration tap the button 9 ...

Page 249: ...urement item and sample type PS and U When a dilution factor is not configured for each sample on the sample information registration screen measurement is performed using the dilution factor config ured here Next page buttons 1 Tap the Data input button NOTE When the button is tapped all dilution factor settings will be printed 2 Tap the button of the measurement item to be edited 9 ...

Page 250: ...n menu 897N201058J FDC NX700 Instruction Manual 2022 04 3 For the sample type liquid type tap a button and then select from D1 D2 or D3 4 For the dilution factor tap a button and then select from None 2 3 4 5 or 10 9 ...

Page 251: ...S D3 D1 CRP Diluent D2 Purified water Saline solution GLU BUN UA TCHO NH3 TG CRE TP ALB TBIL HDLC Ca IP DBIL Mg GGT GOT AST GPT ALT CPK LDH ALP AMYL LAP CKMB CHE CRP NaKCl TCO2 LIP Diluent position for sample type U D1 D2 D3 GLU BUN UA Ca AMYL NaKCl CRP Diluent Purified water Saline solution When an inactivated serum is to be used a saline solution cannot be used NOTE When an inactivated serum is ...

Page 252: ... holder 9 3 16 Print setting for measured result Configure the print quantity 1 to 3 sheets and print order of measurement results Tap the print quantity you want to configure For Print order you can select Slide stacking order or Fixing NOTE If Fixing is selected printing occurs in the following order Plasma serum and urine TP ALB TBIL DBIL GOT AST GPT ALT ALP LDH CHE GGT LAP CPK AMYL TCHO HDLC T...

Page 253: ...ears so you can edit the value IMPORTANT If the entry of the acceptable range is incorrect the judgement of the control measurement results is incorrect Always make sure that the entered values are correct NOTE The units configured by the Unit conver sion of the function menu are displayed NOTE If the acceptable range is changed in formation from before the change is not saved 4 To edit a lot numb...

Page 254: ...urement tap the checkbox to select it NOTE By default the checkbox is cleared not enabled 9 3 19 Operator ID setting Configure whether to display the Operator ID at the Measurement preparation screen and Measure ment results screen NOTE When the USM function is disabled the following setting will appear In case that the USM is enabled refer to Section 11 3 To display the operator ID tap the checkb...

Page 255: ...TE If ALP LDH is selected ALPi LDHi ap pears the same as ALP LDH Check this setting again before testing 3 Choose the item names to be used Tap either the BUN or UREA button NOTE When the name of BUN is changed to UREA or vice versa set again the fol lowing items ITEMS Reference interval Correlation coefficient Control fluid settings NOTE When BUN UREA is changed the change is reflected in the fol...

Page 256: ...on the Function menu screen the pass word input may be required The password can be changed here If the input field is tapped the software keyboard refer to Section 2 3 appears so you can edit the password NOTE A password can be 4 or 5 alphanumeric characters long NOTE In the event of a password being lost please contact your dealer for assistance 9 ...

Page 257: ...o prepare the software and a cable to receive data Before you use this function for the first time please contact your dealer for assistance NOTE Use a shielded LAN cable Otherwise electromagnetic interference may occur NOTE When a LAN or RS 232C cable is not connected measurement results are not sent Make sure a cable is connected NOTE When connecting this analyzer to a network make sure it is pl...

Page 258: ......

Page 259: ...unction menu button is tapped an Operator ID input display is shown Then input Operator ID If input operator ID is registered and also within expiry the meas urement starts or the function menu becomes active NOTE When not registered ID expired ID Emer gency ID counted up to the maximum us age limit are input measurement does not start nor mode function screen become ac tive 11 ...

Page 260: ...erator ID level NOTE OPE Operator ADM Administrator EMG Emergency Measurement Result OPE ADM When expired slide is used EXP is indi cated instead of measurement result EXP is used not only on screen indica tion but also on print and in communi cation This is also the case for CRP measure ment with expired calibration EMG When expired slide is used in EMG measurement result with mark is indi cated ...

Page 261: ...USM User Safety Mode 11 3 897N201058J FDC NX700 Instruction Manual 2022 04 Mode function OPE EMG Operable function is limited to Control Calibration START Manual pipetting and some maintenance functions 11 ...

Page 262: ...11 USM User Safety Mode 11 4 897N201058J FDC NX700 Instruction Manual 2022 04 ADM All functions are operable 11 ...

Page 263: ...ile name dcn_uid_yyyymmddhhmmss dat Ex dcn_uid_20220425201612 dat For the detail see the manual of the Operator ID Editor 2 Create the following folder structure in your USB Rename the file which is created by the tool to dcn_uid dat Store it in the USB WData GUI dcn_uid dat Please use FUJIFILM recommended USB only 3 Select Operator ID setting in ADM mode 4 Select Operator ID import 11 ...

Page 264: ...11 6 897N201058J FDC NX700 Instruction Manual 2022 04 5 The left pop up is shown Then set USB and after a few seconds select OK 6 The left pop up will be shown when im port is complete Take off the USB and then tap Shutdown 11 ...

Page 265: ...11 7 897N201058J FDC NX700 Instruction Manual 2022 04 11 3 2 Operator ID Export 1 Select Operator ID setting in ADM mode 2 Select Operator ID export 3 The left pop up is shown Then set USB and after a few seconds select OK 11 ...

Page 266: ... be shown when ex port is complete Then tap OK and take off the USB 5 Operator ID data dcn_uid dat is stored in below folder structure below in the USB WData GUI dcn_uid dat Browsing and editing of dcn_uid dat is possible in Operator ID Editor For the detail see the manual of the Operator ID Editor 11 ...

Page 267: ...ng registered operator ID Registered Operator ID can be edited as the following procedure NOTE Editable items are the expiry of ID and the level 1 Select Operator ID setting in Mode function for administrator screen 2 Select Operator ID setting 3 Select an Operator ID which needs edit ing 11 ...

Page 268: ...User Safety Mode 11 10 897N201058J FDC NX700 Instruction Manual 2022 04 4 To change the expiry touch the Expira tion field and edit it using the software keyboard 5 To change the level select either ADM OPE 11 ...

Page 269: ...iting emergency ID Dis qualification of registered Emergency ID can be set Also measurement limit count by use of Emergency ID can be set 1 Select Operator ID setting in Mode function for administrator screen 2 Select Edit Emergency ID 3 To dis qualify the Emergency ID tap the checkbox 11 ...

Page 270: ...reset the current count tap Reset button To set a new limit count select either 0 100 or 200 In case of 0 emergency ID does not function NOTE Measurement count by use of Emergency ID cannot exceed the limit count When the current count is up to the limit the measurement by use of Emergency ID is not possible In this case reset the current count 11 ...

Page 271: ...0 005 OD 600 nm Display 7 inch color touch panel Backup data of test results For 270 samples Printer Thermal type paper size 58 mm x 25 m Warming up time Approx 10 minutes 25 C approx 20 minutes 15 C Environmental conditions Indoor use 6 000 cd m2 lux or lower 3 000 cd m2 lux or lower when using the sample barcode reader Altitude Up to 2 000m 802hPa Transient overvoltage category II Pollution degr...

Page 272: ......

Page 273: ...the display and printout messages Abbreviation Meaning AD Analog Digital ADM administrator ave average Bef comp Before compress cali calibration charac character comm communication ctrl control DA Digital Analog Decomp Decompress DL FUJI DRI CHEM DILUENT DL CRP EMG emergency ERR error EXP expiration info information NG No Good NSS Normal Saline Solution OD Optical Density OPE operator Pos position...

Page 274: ......

Page 275: ......

Page 276: ...26 30 NISHIAZABU 2 CHOME MINATO KU TOKYO 106 8620 JAPAN https www fujifilm com EU Balcke Duerr Allee 6 40882 Ratingen Germany Oudenstaart 1 5047 TK Tilburg The Netherlands ...

Reviews: