3-10
3 Measurement
897N201058J FDC NX700 Instruction Manual 2022.04
(4) Registering sample information
Tap a sample rack display to display the sample in-
formation registration screen.
START
STOP
0
3
A
B
D
C
E
MYID001
..0160503010
..0160503009
1
8
2
7
3
4
5
6
Enter or select the sample information (sample num-
ber, sample ID, reference standard, and sample type),
and then tap the [OK] button
7
.
1
Enter the sample number using the software key-
2
Obtain order information from the host PC via a
DMS connection.
3
Enter the sample ID using the software keyboard
4
Select a reference standard from the list.
5
Select a sample type from the list.
6
Select a dilution factor from the list.
Refer to Section 3.6 and Section 9.3.14
7
Enter the registration information, and then re-
turn to the Measurement-preparation screen.
8
Cancel entry of the registration information, and
then return to the Measurement-preparation screen.
NOTE
When the analyzer is connected to DMS, sample information can be acquired automati-
cally from the work list.
IMPORTANT:
The sample type of the sample, slide, and the sample type setting of the FDC NX700 must be same.
Otherwise, an incorrect measurement result will be obtained.
IMPORTANT:
Be sure to enter the correct sample number and sample ID.
IMPORTANT:
When measuring a low concentration protein fluid (such as urine), some bubbles may appear dur
-
ing measurement in the tip. If bubbles have formed during measurement, perform the measure-
ment again from the beginning.
3