12-1
897N201058J FDC NX700 Instruction Manual 2022.04
12
Specifications
Processing speed
190 tests per hour (simultaneous CM and ISE measurements)
Number of incubator cells
CM: 13; ISE: 1
Incubation temperature
37°C (CM) / 30°C (ISE)
Incubation time
1 to 6 minutes
Capacity for simultaneously
loaded slides
20
Slide lot information
Each item: maximum 5 lots
Sampler
(1) Number of samples
5 samples (on the sample disk)
(2) Sample application fluid
volume
50 µl or less (automatic switching)
(3) Dilution
Automatic dilution (maximum 21x dilution)
(4) PF filtering time
Max. 90 sec.
Slide ejection
Automatic ejection (max. 150 slides into the disposal box)
Used tip ejection
Automatic ejection (max. 75 tips into the disposal box)
Measurement wavelengths
400 nm, 415 nm, 505 nm, 540 nm, 577 nm, 600 nm, 625 nm, 650 nm
Light source
Halogen lamp (6 V, 10 W)
Measurement precision
σ
<= 0.0004 OD / 5 minutes
Measurement accuracy
Within ±0.005 OD (600 nm)
Display
7-inch color touch panel
Backup data of test results
For 270 samples
Printer
Thermal type (paper size: 58 mm x 25 m)
Warming up time
Approx. 10 minutes / 25°C, approx. 20 minutes / 15°C
Environmental conditions
Indoor use (6,000 cd/m
2
(lux) or lower. 3,000 cd/m
2
(lux) or lower
when using the sample barcode reader.)
Altitude: Up to 2,000m (802hPa)
Transient overvoltage category: II
Pollution degree: 2
Operating temperature: 15 to 32°C
Operating humidity: 30 to 80% RH (no vapor condensation)
Storage and transport conditions
Temperature: -10 to 50°C
Humidity: 10 to 90% RH (no vapor condensation)
Voltage used
Single phase AC; 100 - 240 V ±10%; 50 to 60 Hz
Consumption current
3 - 1.3 A
Type of protection against electri-
cal shock
CLASS 1 EQUIPMENT
Volume level
Within 70 dB (including buzzer sound)
Data transfer
RS-232C (1 port); USB (1 port); LAN (1 port)
External dimensions
(W) 500× (D) 380× (H) 410 mm
Weight
33kg
Effective period of service
(service life)
6 years after installation
(This period is valid as long as the precautions for use are followed
and regular periodic maintenance is performed correctly.)
12