3-9
3 Measurement
897N201058J FDC NX700 Instruction Manual 2022.04
Slide cartridge
4. Align the direction and load the slide
cartridge at the sample location to be
measured.
0 ml
5
.
φ
5
13
×
7
0.5ml
13
75
×
5. Make sure the sample rack and sample
tube types are correct.
NOTE
To avoid a suction error or malfunction,
use a sample tube that matches the la-
bel on the sample rack.
NOTE
Refer to Section 3.9 for sample tubes
that can be used.
Blood collecting tube
or FUJI Sample Tube
6. Remove the cap of the blood collecting
tube or FUJI Sample Tube that contains
sample, and then load it on the dedicat-
ed sample rack.
NOTE
Make sure there are no bubbles in the
sample in the blood collecting tube or
FUJI Sample Tube. Use a centrifuge to
remove bubbles. If bubbles are present,
the sample may be detected incorrectly.
IMPORTANT:
Be sure not to mistake the positions of the sample
and its measurement slides.
A
B
C
D
E
Sample rack
Slide cartridge
NOTE
When registering one sample informa-
tion to multiple sample racks (A to E)
for testing, set the sample and slides at
each sample position.
NOTE
Even when registering one sample in-
formation to multiple sample racks (A to
E) for testing, be sure to load the slides
used for calculation items in the same
slide cartridge to perform the measure-
ments.
3