6-38
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
PF pump
Joint (G)
4. While pushing the head of joint (G), pull
the joint (G) to the right to remove the
tube (D).
Pull off joint (G) to the right
while pushing this part to
the right.
Red tube (D)
Joint (G)
Holder (C)
Tube (H)
5. Slide the holder (C) to the left and load
the new tube (H) on to the rotor.
NOTE
As shown in the diagram to the left,
align the base of the joint (E) with the
cover hole (F).
NOTE
Set so that the two front marks face the
front.
Front Marks
Cover hole (F) Joint (E)
Joint (E)
Cover hole (F)
When attaching the cover (B), make
sure that the hooks are hooked and
interlocked.
NOTE
If attachment is incorrect, a “W0134 PF
sampling error” may occur.
6. Reattach the cover (B) using the screw
(A).
Joint (G)
Red tube (H)
Line
7. Push in the new tube (H) strongly until
the line of the joint (G).
NOTE
Insert the tip of the red tube (H) in the
joint (G) until the line.
6