9-12
9 Function menu
897N201058J FDC NX700 Instruction Manual 2022.04
•
If [Serial connection] is set, configure the com
-
munication type and whether to output from the
printer. You can choose from 4 communication
types.
- Type 1: Bidirectional connection
- Type 2: Unidirectional connection (bidirectional
format)
- Type 3: Unidirectional connection (standard
format)
NOTE
Rebooting the device is necessary for
the setting to take effect. The confirma
-
tion dialog for the power-off will appear
when
is tapped to exit Mode function
for administrator. After the shutdown, re-
boot the analyzer.
•
If [LAN connection] is set, configure the com
-
munication type and whether to output from the
printer. You can choose from 4 communication
types.
- Type 1: Bidirectional connection
- Type 2: Unidirectional connection (bidirectional
format)
- Type 3: Unidirectional connection (standard
format)
- Type 4: Unidirectional connection (HL7 for-
matted)
• When
was tapped, the LAN configuration
screen appears. If an entry field is tapped, the
software keyboard (refer to Section 2.3) appears
so you can edit the value.
NOTE
The following section shows the default
LAN settings.
FDC NX700 settings
IP address:192.168.100.1
Subnet mask:255.255.255.0
Port number:65535
Gateway: 0.0.0.0
PC settings
IP address:192.168.100.2
Port number:65535
NOTE
Rebooting the device is necessary for
the setting to take effect. The confirma
-
tion dialog for the power-off will appear
when
is tapped to exit Mode function
for administrator. After the shutdown, re-
boot the analyzer.
9