8 Troubleshooting
8-12
897N201058J FDC NX700 Instruction Manual 2022.04
(2) When tip cannot be ejected
Error
code
Error message
Error condition
Troubleshooting
E0086
Sample tip eject ERR
After analyzer shut-
down, empty disposal
box, and then turn the
power ON.
Tips cannot be dis-
carded.
After analyzer shutdown, empty dis-
posal box, and then turn the power ON.
E0087
Reference fluid tip
eject ERR
After analyzer shut-
down, empty disposal
box, and then turn the
power ON.
W5401
Reconfirmation
An error occurred while
discarding the tip at
the last startup.
Please check the dis-
posal box.
If the disposal box is
not empty, please shut
down, empty the dis-
posal box, and restart.
Please check the disposal box.
If the disposal box is not empty, please
shut down, empty the disposal box, and
restart.
W5402
Reconfirmation
An error occurred while
discarding the tip at
the last startup.
Please check that the
sample application unit
cover is set.
If the sample applica-
tion unit cover is not
set, please shut down,
assemble the sample
application unit cover,
and restart.
Please check that the sample applica-
tion unit cover is set.
If the sample application unit cover is
not set, please shut down, assemble
the sample application unit cover, and
restart.
8