5.2 Setting tips
Tips are required for all measurements. Check the number of remaining tips on the LCD before per-
forming measurements. If necessary, add new tips.
WARNING:
If the analyzer is applying samples, tap the [PAUSE] button to interrupt application (refer to Sec-
tion 4.2), and then load the tips. It is unsafe to set tips during sample application due to the mov-
ing parts of the sampler unit and the sample disk.
Tips
1. With [Warming up], [Lamp warming up]
or [Ready to test] displayed on the LCD,
make sure that the analyzer is not apply-
ing samples.
2. Tap the tip display on the LCD.
3. The Tip loading screen appears.
The sample disk turns, and the consumables
are moved to the front.
CAUTION:
Do not touch the sample disk until it stops.
5