9-14
9 Function menu
897N201058J FDC NX700 Instruction Manual 2022.04
9.3.3 Calculated Parameter
You can configure the settings to calculate the following items from simultaneously measured values
and then display the results. Items registered for calculation have their measurement results dis-
played and printed.
Test name
Calculation
test name
Unit
Calculation formula
LDL Cholesterol (LDL-C)
LDL
mg/dL
or
mmol/L
LDL-C = TCHO value - (HDL-C value + TG value/5) [mg/dL]
LDL-C = TCHO value - (HDL-C value + TG value/2.2) [mmol/L]
NOTE
LDL-C is NOT calculated when the measured value of
TG exceeds 400 mg/dL (4.52 mmol/L).
non-HDL Cholesterol
(non HDL-C)
non-HDL
mg/dL
or
mmol/L
non-HDL = TCHO value - HDL-C value
Globulin
GLOB
g/dL or
g/L
GLOB = TP value - ALB value
Albumin/Globulin ratio
ALB/GLOB
—
ALB/GLOB = ALB value / (TP value - ALB value)
BUN/Creatinine ratio
BUN/CRE
mg/mg
or
(SI)
BUN/CRE = BUN value/CRE value [mg/mg]
BUN/CRE = BUN value ×1,000/CRE value[(SI)]
NOTE
BUN measured value in SI unit (mmol/L) is
shown as molar concentration of Urea. BUN/
CRE ratio in SI unit (molar ratio) is shown Urea
concentration/creatinine concentration ratio.
GOT/GPT ratio
(AST/ALT ratio)
GOT/GPT
(AST/ALT)
—
GOT/GPT = GOT value / GPT value
(AST/ALT = AST value / ALT value)
NOTE
The calculation test name can be changed by
[Test item name setting] in the Mode function for
administrator.
Sodium/Potassium ratio
Na/K
—
Na/K = Na value/K value
Anion Gap
Anion Gap
mEq/L
or
mmol/L
Anion Gap = Na value - (Cl value + TCO
2
value)
IMPORTANT:
Calculation items display a value calculated from simultaneously measured values. Therefore, cal-
culation results are impacted by errors included in measurement values and the effects of interfer
-
ing materials. Please understand that when you use the results for a diagnosis, you must make an
overall determination in combination with the measurement values is used in the calculation.
NOTE
In the case of additional testing and retesting, items subject to calculation must be meas-
ured at the same time.
NOTE
When there are two or more slides of the same item subject to calculation in the slides
measured at the same time, the initial measurement value is used.
NOTE
When the measurement results of items subject to calculation exceed the measurement
range (determination range), the calculation results of this calculation item are displayed
and printed as "****".
NOTE
When an item subject to calculation has expired, an (expiry) warning symbol (#) is displayed
and printed for the measurement results of this item and the calculation results.
9