4-18
4 Applicable usage methods
897N201058J FDC NX700 Instruction Manual 2022.04
Spot now
Stop spottin
Preparing manual spotting
START
9. After the progress bar has finished (all
□ change to ■), apply the sample to the
slide while [Spot now] is displayed.
The spotted slide is automatically transferred
to the measurement section after the display of
[Spot now] stops.
IMPORTANT:
Apply the sample while the buzzer is sounding
(while [Spot now] is displayed). Otherwise, the
measurement results will be inaccurate.
IMPORTANT:
When applying sample to a CM slide, apply the sam-
ple within 1 mm of the center of the slide. Otherwise,
the measurement results will be inaccurate. Run the
measurement again.
参照液
検体
搬送方向
(インキュベータ方向)
スライドの中央に点着する
(比色スライド)
Apply sample to the center
of the slide (CM slide).
NOTE
The LCD may not display [Wait] de-
pending on the operation conditions.
10. To apply sample to the next slide, repeat
the procedure from Step 5.
11. Press the [HOME] button to end manual
sample application.
NOTE
The sampler unit moves.