6-15
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
6.3.3 Reassembling parts and ending cleaning
5
2
3
4
1
Thumbscrew
1. Reassemble the incubator.
Align the arrow marks on the incubator and the
reference plate, and then tighten the 5 thumb-
screws evenly by tightening opposing thumb-
screws in numerical order (
1
to
5
).
IMPORTANT:
Tighten the five thumbscrews securely. Otherwise,
the measurement results may be negatively af-
fected.
1
Slide transfer
weight
5
Panel unit
pedestal
2
Sample application unit
3
Incubator cover
4
Incubator cable
Incubator cover
Incubator cable
2. Reassemble each part as shown in the
illustration on the left.
1
Load the slide transfer weights.
2
Load the sample application unit cover
and tighten the thumbscrews.
3
Load the incubator cover.
4
While aligned with the groove, plug the
connector of the incubator cable into the
incubator and then lock it by turning it to
the right.
5
Load the panel unit pedestal.
IMPORTANT:
Tighten the thumbscrews securely. Otherwise, the
measurement results may be negatively affected.
IMPORTANT:
Be sure to attach the incubator cover. Otherwise,
the measurement results may be negatively af-
fected.
NOTE
When you insert the connector, make
sure not to twist or apply too much pres-
sure to the cable. If the cable is twisted
more than once, it may break.
3. Close the panel section to end cleaning.
Run [Reference plate level check] again, and
then check the results.
NOTE
Wait until the temperature of the incuba-
tor is steady.
6