6.6 Inspecting and Replacing the Sampler O-ring
The sampler nozzle O-ring wears over time. Inspect it once a month and replace it once a year.
WARNING:
When inspecting and replacing O-rings, always wear gloves, a lab coat, and safety goggles.
If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un-
der running water, and then disinfect the area.
Seek medical assistance if necessary.
IMPORTANT:
If you do not inspect and replace the sampler O-ring, the sample application volume will be inac-
curate, which may negatively affect the measurement values.
6.6.1 Inspecting the sampler O-ring
1. Tap the button.
2. Tap the [Normal mode] button.
6