SONET/SDH Configuration and Testing
Setting Up for Testing STS-3c and STM-1 Signals
4
4-11
Changing the Threshold Setting for Counters
The
Err Thresh
option on the
STS-3c
or
STM-1 SETUP
submenu is used to specify
the number of errors or events that must occur in one second for the threshold
counter to increment by one. The default setting for this option is 30.
Note
The sample screens presented in this procedure are STS-3c screens.
If the framing format you selected is STM-1, the OC3port Plus
displays equivalent STM-1 screens.
To change the threshold setting for counters, do the following:
1.
Press the
SONET
key. The
SONET SELECTIONS
menu is displayed.
2.
Press the
Down Arrow
key to move the
Setup STS-3c
or
Setup STM-1
option to
the selection line. Then, press
ENTER
. The
SETUP
submenu is displayed.
3.
Press the
Down Arrow
key to move the
Err Thresh
option to the selection line.
For example:
STS-3c SETUP:
S-Bit Value 00
Err Thresh (30)
Path FERF Suprt ON
za045f.eps
4.
Press
ENTER
. You see this display:
ERROR THRESHOLD
THRESHLD CNT:000030
za046f.eps
You enter Edit mode as indicated by the flashing number in the first position
of the
THRESHLD CNT
field.