Testing ATM Networks
Quality of Service Tests
6
6-63
5.
Make sure that the
Duration
parameter is on the selection line.
This parameter determines how long the slave unit transmits the burst of cells
to the master unit. The default setting is
005.0
seconds.
6.
Press the
EDIT
key and enter the same value for Duration that you specified
when you set up the master unit. Press
ENTER
when you finish to save the
change.
7.
Press the
Down Arrow
key to move the
RT Delay
parameter to the selection line.
This parameter determines the amount of time the master unit waits to receive
the cells transmitted from the slave unit.
Note
The amount of time the master unit waits begins when it sends the
initial trigger to the slave unit and ends when it receives the cells
sent from the slave unit.
When setting the
RT Delay
value, you need to take into consideration
the time for the initial transmission (the trigger), the slave unit’s
receipt of the trigger, and the slave unit’s transmission of cells back
to the master unit. Because of inherent delays in this process, you
will probably have to increase (as compared with a regular Cell Loss
test) the value of the
RT Delay
parameter.
8.
Press the
EDIT
key and enter the same value for the
RT Delay
parameter that
you specified when you set up the master unit.
9.
Press
ENTER
when you finish to save the change.
You have completed the setup for the Remote Cell Loss test.