OC3port Plus
Users Manual
3-16
1.
Press the
OPTIONS
key. The
OPTION SELECTION
menu is displayed:
OPTION SELECTIONS
- - - - - - - - - - - - - - - - - - - - - - -
Parameter Set 2
RS-232 Speed 96
za019f.eps
2. Press the
Down Arrow
key until the
Factory Defaults
option is on the selection
line:
3. Press
ENTER
. The unit issues a short "beep" and resets to the factory defaults.
You return to the introductory display.
Clearing the Error Counters
If there are errors on a signal, they are presented on the LCD. As certain errors are
detected, the counters on the display continuously increase. To clear the display of
any errors and reset the error counter to zero, press the
CLEAR
key while you are
viewing the errors on the LCD.
Responding to Low Battery Warnings
The OC3port Plus uses NiCad internal batteries. These batteries last for about one
year. When the battery charge is low, the OC3port Plus issues periodic "beeps"
and then displays a
LOW BATTERY
warning on the LCD. When this situation
occurs, the unit has approximately five minutes of battery power remaining. We
suggest that you use the remaining time to immediately store any current test
results and then recharge the unit.
Note
If the battery does not charge, contact Fluke Customer Service for
replacement batteries.
You can connect the tester to the supplied AC adapter/charger and charge it
without shutting it off. On a full battery charge, which takes between 12 to 14
hours, the OC3port Plus will run for approximately three hours.
Caution
The AC adapter/charger should never be disconnected
while the OC3port Plus is powered on. This can cause
corruption of current or stored test results.