OC3port Plus
User Manual
6-58
4.
Press the
Down Arrow
key to move the
Setup Cell Loss
option to the selection
line. Then, press
ENTER
. The
CELL LOSS PARAMETERS
submenu is
displayed:
CELL LOSS PARAMETERS
- - - - - - - - - - - - - - - - - - - - - - - -
Duration: 005.0
RT Delay: 0.05
za136f.eps
5. Make sure that the
Duration
parameter is on the selection line. This parameter
determines how long the slave unit transmits the burst of cells to the master
unit. The default setting is
005.0
seconds.
Press the
EDIT
key and enter a value for the Duration parameter. Press
ENTER
when you finish to save the change.
6. Press the
Down Arrow
key to move the
RT Delay
parameter to the selection line.
This parameter determines the amount of time the master unit waits to receive
cells back from the slave unit.
Notes
The amount of time the master unit waits begins when it sends the
initial trigger to the slave unit and ends when it receives the cells
sent from the slave unit.
When setting the
RT Delay
value, you need to take into consideration
the time for the initial transmission (the trigger), the slave unit’s
receipt of the trigger, and the slave unit’s transmission of cells back
to the master unit. Because of inherent delays in this process, you
will probably have to increase (as compared with a regular Cell Loss
test) the value of the
RT Delay
parameter.
7. Press the
EDIT
key and enter a value for the
RT Delay
parameter. Press
ENTER
when you finish to save the change.
You have completed the configuration of the master unit for the
Remote Cell
Loss
test.