![Fluke 750SW User Manual Download Page 216](http://html1.mh-extra.com/html/fluke/750sw/750sw_user-manual_2310283216.webp)
DPCTrack2
Users Manual
A-2
other calibration report, but instead of summarizing the stated accuracy, accuracy
percentage and resolution at the top of the report, it displays this information for each test
point individually.
Instrument Test Point Analysis
This report shows an analysis of a single Instrument calibration. A line graph is included
so that you may easily compare the expected, As-Found, and As-Left values for the
calibration.
Instrument Test Point Trend
This report allows you to trend an Instrument’s maximum deviation over a specified
period of time. A line graph is included so that you may easily view this information.
Loop Calibration Report (Spec Summary)
This is a completed calibration report for Loops showing some master information,
results of the calibration (calibration specs plus “As-Found” and “As-Left” values and
deviation percentage), its comprising Instrument(s), and the Test Instrument(s) used. This
is a good report to choose for those Loops whose calibrations are fairly simple- those
with test points that remain consistent with regard to stated accuracy, accuracy
percentage and resolution since this information is summarized at the top of the report.
Loop Calibration Report (Specs/Test Point)
This is also a completed calibration report for Loops, although this report shows more
detail than the one previously described. This is a good report to choose for those Loops
whose calibrations are more complex. It includes the same information as the other Loop
calibration report, but instead of summarizing the stated accuracy, accuracy percentage
and resolution at the top of the report, it displays this information for each test point
individually
Loop Test Point Analysis
This report shows an analysis of a single Loop calibration. A line graph is included so
that you may easily compare the expected, As-Found and As-Left values for the
calibration.
Loop Test Point Trend
This report allows you to trend a Loop’s maximum deviation over a specified period of
time. A line graph is included so that you may easily view this information.
Test Instrument Calibration (Spec Summary)
This is a completed calibration report for Test Instruments showing some master
information, results of the calibration (calibration specs plus “As-Found” and “As-Left”
values and deviation percentage), and the Test Instrument(s) used. This is a good report
to choose for those Test Instruments whose calibrations are fairly simple- those with test
points that remain consistent with regard to stated accuracy, accuracy percentage and
resolution since this information is summarized at the top of the report.
Test Instrument Calibration (Specs/Test Point)
This is also a completed calibration report for Test Instruments, although this report
shows more detail than the one previously described. This is a good report to choose for
those Test Instruments whose calibrations are more complex. It includes the same
information as the other Test Instrument calibration report, but instead of summarizing
the stated accuracy, accuracy percentage and resolution at the top of the report, it displays
this information for each test point individually.