DPCTrack2
Users Manual
18-14
Authorize Calibration Screen. See Test Instrument Recognition in Chapter 18.9 for
more information.
6.
Process all test points and readings according to the item's default test point groups if
found, or it will generate default information for those test points and groups that are
not found in the database.
7.
Build a new Calibration record and display the Authorize Calibration Screen, for
verification of all processed and uploaded data.
8.
If authorized, the program will add the displayed record as a new, unfinalized
Calibration record. To finalize or check over your calibration records, start
DPCTrack2, log in with your user name and password, and open the Calibration
Record Screen.
Note
When the Upload is completed, you may be asked if you want to clear the
calibrator memory. (This is to prevent re-upload of calibration results.) See
the Warnings below if you choose to clear the memory at this point:
Cauton
•
Incomplete tasks will be cleared from memory in this
process. Keep track of any instruments you may need to
re-download.
•
Non-uploadable results will be cleared from memory in
this process. Record all data that you wish to keep
separately.
•
If result data is not cleared, it may be unintentionally
uploaded again, and duplicate calibration records may
result. This is why it is recommended that you clear the
calibrator of data following a successful Upload.
18.8.4.1 Result Duplicates
The upload function of the program only allows for one new calibration record for a
particular Instrument or Loop during one Upload session.
If multiple results are Uploaded for a particular tag, the utility will display the date and
time (if available) of the result record and ask the user to choose the one group that
should be added to the current calibration record.
If you need to upload both cal records, perform the upload twice, choosing different
groups each time the Group Result Duplicate screen appears.