DPCTrack2
Users Manual
15-6
To begin adding this information, you must first click the Add Group button, which will
bring up blank calibration specification fields and a blank grid. Your screen will look
similar to that in Figure 15.2 just without any data. You may enter any
Group Name
you
like (up to 50 characters) and continue through the rest of the fields. The information
entered for each group will remain unique to that group. As you enter information into
each field and hit the tab key, the values you have entered or the selections you have
made will automatically be inserted into the calibration specifications grid below. The
number of rows in the grid is determined by the number of test points you enter into the
#
of Cal Pts
field.
15.2.2.1 Test Type
For standard input-output tests, leave this field blank. Two additional test types are
provided and are needed, described below.
15.2.2.1.1 Manual Pass/Fail
After the Test Type Package is selected in the Test Type field on the Calibration Points
tab of the instrument record, the screen refreshes to display the Manual Pass/Fail format
as shown in Figure 15.3. This can be used to capture the Pass or Fail of any test without
measurement data, either via Calibration record entry or via DPC download.
Test Inputs:
The inputs can be set to default values to appear when entering a
Calibration record, they can also be changed during a Calibration event.
•
As Found Result
o
Pass
– Select Pass if there are no As Found out of tolerances.
o
Fail
– Select Fail if there are As Found out of tolerances.
o
Unknown
– Select Unknown if As Found vales are not applicable.
o
Blank
– Leaving the field blank will default to “Not Entered” which will
have the same result as “Unknown”.
•
As Left Result
o
Pass
– Select Pass if there are no As Left out of tolerances.
o
Fail
– Select Fail if there are As Left out of tolerances.
o
Unknown
- Select Unknown if As Left vales are not applicable.
o
Not Entered
– Select Not Entered if As Left values were not recorded.
o
Blank
– leaving this field blank will result in the Pass/Fail logic to be
determined by the As Found Result.
•
Test Performed
o
No
– If this field is set to ‘No’ or is left Blank, no failure fields will be set
regardless of other data entered for this Group. The Calibration can be saved
as either Passed or Failed
.
o
Yes
– If set to ‘Yes’ it will use the information in the As Found and As Left
fields to determine the Pass/Fail results.