Operation options
Proline Promass F 200 HART
32
Hauser
8.2.2
Operating philosophy
The individual parts of the operating menu are assigned to certain user roles. Each user role
corresponds to typical tasks within the device lifecycle.
Menu
User role and tasks
Content/meaning
Language
task-oriented
Role "Operator", "Maintenance"
Tasks during operation:
• Configuring the operational display
• Reading measured values
Defining the operating language
Display/operat.
• Configuring the operational display (e.g. display format,
display contrast)
• Resetting and controlling totalizers
Setup
"Maintenance" role
Commissioning:
• Configuration of the measurement
• Configuration of the outputs
Wizards for quick commissioning:
• Defining the medium
• Configuring the outputs
• Configuring the operational display
• Defining the output conditioning
• Configuring the low flow cut off
• Configuring the monitoring of partial and empty pipe
detection
"Advanced setup" submenu:
• For more customized configuration of the measurement
(adaptation to special measuring conditions)
• Configuration of totalizers
Diagnostics
"Maintenance" role
Fault elimination:
• Diagnostics and elimination of process and device
errors
• Measured value simulation
Contains all parameters for error detection and analyzing
process and device errors:
•
"Diagnostic list" submenu
Contains up to 5 currently pending diagnostic messages.
•
"Event logbook" submenu
Contains up to 20 or 100 (order option) event messages that
have occurred.
•
"Device information" submenu
Contains information for identifying the device.
•
"Measured values" submenu
Contains all current measured values.
•
"Data logging" submenu (order option)
Storage and visualization of up to 1000 measured values
•
"Simulation" submenu
Is used to simulate measured values or output values.
•
"Device reset" submenu
Resets the device configuration to certain settings
Expert
function-oriented
Tasks that require detailed knowledge of the function
of the device:
• Commissioning measurements under difficult
conditions
• Optimal adaptation of the measurement to difficult
conditions
• Detailed configuration of the communication
interface
• Error diagnostics in difficult cases
Contains all the parameters of the device and makes it possible
to access these parameters directly using an access code. The
structure of this menu is based on the function blocks of the
device:
•
"System" submenu
Contains all higher-order device parameters that do not
pertain either to measurement or the measured value
communication.
•
"Sensor" submenu
Contains all parameters for configuring the measurement.
•
"Output" submenu
Contains all parameters for configuring the analog current
outputs.
•
"Communication" submenu
Contains all parameters for configuring the digital
communication interface.
•
"Application" submenu
Contains all parameters for configuring the functions that
go beyond the actual measurement (e.g. totalizer).
•
"Diagnostics" submenu
Contains all parameters for error detection and analyzing
process and device errors and for device simulation.