Proline Promass F 200 HART
Operation
Hauser
93
Preset value 3
Reset all totalizers
Parameter overview with brief description
Parameters
Prerequisite
Description
Selection/
User entry
Factory setting
Control totalizer 1-3 One of the following options is
selected in the
Assign process
variable
parameter of the
Totalizer 1-3
submenu:
• Mass flow
• Volume flow
• Corrected volume flow
Control totalizer
value.
• Totalize
• Reset + hold
• hold
• Reset + totalize
• totalize
Totalize
Preset value 1-3
One of the following options is
selected in the
Assign process
variable
parameter of the
Totalizer 1-3
submenu:
• Mass flow
• Volume flow
• Corrected volume flow
Specify start value
for totalizer.
Floating point
number with sign
Country-dependent:
• 0 kg
• 0 lb
Reset all totalizers
-
Reset all totalizers to
0 and start.
• Cancel
• Reset + totalize
Cancel
11.7
Showing data logging
In the device, the extended function of the HistoROM must be enabled (order option) so that the
Data logging
submenu appears. This contains all parameters for the measured value history.
Navigation path
"Diagnostics" menu
®
Data logging
Function scope
• A total of 1000 measured values can be stored
• 4 logging channels
• Adjustable logging interval for data logging
• Display of the measured value trend for each logging channel in the form of a chart
A0016357
å
24
Chart of a measured value trend
• x-axis: depending on the number of channels selected displays 250 to 1000 measured values of
a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the ongoing
measurement.