Commissioning
Proline Promass F 200 HART
62
Hauser
Sequence of the wizard for the frequency output
Val. at min.freq
End of wizard
Invert outp.sig.
Volume flow
Off
Min. freq. value
Max. freq. value
Operating
mode
Frequency
Val. at max.freq
Failure mode
Density
Mass flow
Correct.vol.flow
Ref.density
Temperature
Volume flow
unit
Density unit
Mass flow
unit
Corr. vol. unit
Temperature
unit
Actual value
Defined value
0 Hz
Failure freq.
Assign freq.
Ref. dens. unit
A0018171-EN
å
15
"Pulse/frequency/switch output " wizard in the "Setup" menu: "Frequency" operating mode
Parameter overview with brief description for frequency output
Parameter
Description
Selection/
User entry
Factory setting
Operating mode
Specify the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
The picklist depends
on the selected device
order (
Pulse