Diagnostics and troubleshooting
Proline Promass F 200 HART
96
Hauser
12
Diagnostics and troubleshooting
12.1
General troubleshooting
For local display
Problem
Possible cause
Remedy
Local display dark and no output signals Supply voltage does not match that
specified on the nameplate.
Apply the correct supply voltage
(
Local display dark and no output signals Supply voltage has incorrect polarity.
Reverse polarity of supply voltage.
Local display dark and no output signals No contact between connecting cables
and terminals.
Check the contacting of the cables and
correct if necessary.
Local display dark and no output signals Terminals are not plugged into the I/O
electronics module correctly.
Check terminals.
Local display dark and no output signals I/O electronics module is defective.
Local display is dark, but signal output
is within the valid range
Display is set too bright or too dark.
• Set the display brighter by
simultaneously pressing
O
+
F
.
• Set the display darker by
simultaneously pressing
S
+
F
.
Local display is dark, but signal output
is within the valid range
Spiral cable of the display module is not
plugged in correctly.
Insert the plug correctly into the main
electronics module and display module.
Local display is dark, but signal output
is within the valid range
Display module is defective.
Backlighting of local display is red
Diagnostic event with "Alarm"
diagnostic behavior has occurred.
Text on local display appears in a foreign
language and cannot be understood.
Incorrect operating language is
configured.
1. Press
S
+
O
for 2 s ("home position").
2. Press
F
.
3. Set the desired language in the
Language
parameter.
For output signals
Problem
Possible cause
Remedy
Signal output outside the valid range
Main electronics module is defective.
Signal output outside the valid current
range
(
<
3.6 mA or
>
22 mA)
I/O electronics module is defective.
Device shows correct value on local
display, but signal output is incorrect,
though in the valid range.
Configuration error
Check and correct parameter
configuration.
Device measures incorrectly.
Configuration error or device is
operated outside the application.
1. Check and correct parameter
configuration.
2. Observe limit values specified in the
"Technical Data".
For access
Problem
Possible cause
Remedy
No write access to parameters
Hardware write protection enabled
Set the write protection switch on the
main electronics module to the OFF
position (