Commissioning
Proline Promass F 200 HART
84
Hauser
10.7
Simulation
The
Simulation
submenu enables you to simulate, without a real flow situation, various process
variables in the process and the device alarm mode and to verify downstream signal chains (switching
valves or closed-control loops).
Navigation path
"Diagnostics" menu
®
Simulation
Structure of the submenu
Simulation
®
Assign simulation process
variable
Value process variable
Simulation current output 1
Value current output 1
Frequency simulation
Frequency value
Pulse simulation
Pulse value
Switch output simulation
Switch status
Simulation current output 2
Value current output 2
Simulation device alarm
Parameter overview with brief description
Parameters
Prerequisite
Description
Selection/
User entry
Factory setting
Assign simulation
process variable
-
Select a process
variable for the
simulation process
that is activated.
• Off
• Mass flow
• Volume flow
• Corrected
volume flow
• Density
• Reference density
• Temperature
Off
Value process
variable
One of the following options is
selected in the
Assign
simulation process variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
Enter the simulation
value for the
selected process
variable.
Depends on the
process variable
selected
-