Proline Promass F 200 HART
Technical data
Hauser
127
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
• In most applications, 20 to 50
%
of the maximum full scale value can be considered ideal
• Select a lower full scale value for abrasive substances (such as liquids with entrained solids): flow
velocity
<
1 m/s (
<
3 ft/s).
• For gas measurement the following rules apply:
– The flow velocity in the measuring tubes should not exceed half the sonic velocity (0.5 Mach).
– The maximum mass flow depends on the density of the gas: formula (
Pressure loss
To calculate the pressure loss, use the
Applicator
17.10 Mechanical construction
Design, dimensions
For the dimensions and installation lengths of the device, see the "Technical Information"
document, "Mechanical construction" section
Weight
Compact version
Weight in SI units
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg]
DN
[mm]
Weight [kg]
8
9
15
10
25
12
40
17
50
28
Weight in US units
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [lbs]
DN
[in]
Weight [lbs]
3/8
20
½
22
1
26
1½
37
2
62
Materials
Transmitter housing
• Order code for "Housing", option
C
: aluminum coating AlSi10Mg
• Order code for "Housing", option
B
: stainless steel 1.4404/316L
• Window material: glass