Mounting
Proline Promass F 200 HART
22
Hauser
DN
E
F
G
H
L
[mm]
[in]
[mm]
[in]
[in]
[in]
[mm]
[in]
[mm]
[in]
40
1½
Approx.42
Approx.1.65
AF 1
½ NPT
67
2.64
310
12.20
50
2
Approx.42
Approx.1.65
AF 1
½ NPT
79
3.11
452
17.78
!
WARNING
Limited functional reliability of the rupture disk.
Danger to persons from escaping fluids!
►
Do not remove the rupture disk.
►
When using a rupture disk, do not use a heating jacket.
►
Make sure that the function and operation of the rupture disk is not impeded through the
installation of the device.
►
Take precautions to prevent damage and danger to persons if the rupture disk is actuated.
►
Observe information on the rupture disk sticker.
6.2
Mounting the measuring device
6.2.1
Required tools
For transmitter
• For turning the transmitter housing: Open-ended wrench8 mm
• For opening the securing clamps: Allen key3 mm
For sensor
For flanges and other process connections: Corresponding mounting tools
6.2.2
Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove any protective covers or protective caps present from the sensor.
3. Remove stick-on label on the electronics compartment cover.
6.2.3
Mounting the measuring device
!
WARNING
Danger due to improper process sealing!
►
Ensure that the inside diameters of the gaskets are greater than or equal to that of the process
connections and piping.
►
Ensure that the gaskets are clean and undamaged.
►
Install the gaskets correctly.
1. Ensure that the direction of the arrow on the nameplate of the sensor matches the flow direction
of the fluid.
2. Install the measuring device or turn the transmitter housing so that the cable entries do not
point upwards.