Proline Promass F 200 HART
Diagnostics and troubleshooting
Hauser
105
835
Process temperature
Increase process temperature.
S
Warning
*
842
Process limit
Low flow cut off active!
1. Check low flow cut off configuration.
S
Logbook entry only
862
Partly filled pipe
1. Check for gas in process.
2. Adjust detection limits.
S
Warning
*
910
Measuring tube does not
vibrate
1. Check process conditions.
2. Increase voltage.
3. Check main electronic module or
sensor.
F
Alarm
912
Inhomogeneous
Medium is inhomogeneous, e.g. gas or
solid content
1. Check process conditions.
2. Increase system pressure.
In particular with outgassing
media and/or increased gas
content, the following measures
are recommended to increase
system pressure:
• Install the instrument at the
outlet side of a pump.
• Install the instrument at the
lowest point of an ascending
pipeline.
• Install a flow restriction, e.g.
reducer or orifice plate,
downstream from the
instrument.
S
Warning
*
913
Inhomogeneous
Oscillation amplitude limit!
The medium properties do not allow a
precise measurement.
Cause: Process medium is very
inhomogeneous (gas or solid content)
1. Check process conditions.
2. Increase voltage.
3. Check main electronic module or
sensor.
S
Warning
*
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section (
12.6
Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the previous
diagnostic event separately.
Navigation path
• "Diagnostics" menu
®
Actual diagnostics
• "Diagnostics" menu
®
Previous diagnostics
Parameter overview with brief description
Parameter
Prerequisite
Description
Display
Actual diagnostics
1 diagnostic event has
occurred
Displays the current
diagnostic event along with
the diagnostic information.
If two or more
messages occur
simultaneously, the
message with the
highest priority is
shown on the display.
Symbol for diagnostic
behavior, diagnostic code
and short message