Diagnostics and troubleshooting
Proline Promass F 200 HART
104
Hauser
411
Up-/download active
Up-/download active, please wait
C
Warning
431
Trim
Carry out trim.
C
Warning
437
Incompatible
configuration
1. Restart device.
2. Contact service.
F
Alarm
438
Dataset
1. Check data set file.
2. Check device configuration.
3. Up- and download new
configuration.
M
Warning
441
Current output 1-2
1. Check process.
2. Check current output settings.
S
Warning
*
442
Frequency output 1
1. Check process.
2. Check frequency output setting.
S
Warning
*
443
Pulse output 1
1. Check process.
2. Check pulse output setting.
S
Warning
*
453
Flow override
Deactivate flow override.
C
Warning
484
Simulation failsafe mode Deactivate simulation.
C
Alarm
485
Simulation process
variable
Deactivate simulation.
C
Warning
491
Simulation current
output 1-2
Deactivate simulation.
C
Warning
492
Frequency simulation 1
Switch off frequency output simulation. C
Warning
493
Pulse simulation 1
Switch off pulse output simulation.
C
Warning
494
Switch output simulation
1
Switch off switch output simulation.
C
Warning
*
Diagnostic behavior can be changed: "Adapting the diagnostic behavior" section (
Diagnostic
number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the process
801
Supply voltage too low
Increase supply voltage.
S
Warning
*
803
Current loop 1-2
1. Check wiring.
2. Change I/O module.
M
Warning
830
Ambient temperature
range
Reduce the ambient temperature
around the sensor housing.
S
Warning
*
831
Ambient temperature
range
Increase the ambient temperature
around the sensor housing.
S
Warning
*
832
Ambient temperature
range
Reduce ambient temperature.
S
Warning
*
833
Ambient temperature
range
Increase ambient temperature.
S
Warning
*
834
Process temperature
Reduce process temperature.
S
Warning
*