Proline Promass F 200 HART
Commissioning
Hauser
73
10.4.7 Configuring the partial filled pipe detection
The
Partial filled pipe detection
wizard guides you systematically through all parameters that
have to be set for configuring the monitoring of the pipe filling.
Navigation path
"Setup" menu
®
Partial filled pipe detection
Sequence of the wizard
Assign process
variable
Off
Density
Reference
density
High value
partial filled
pipe detection
Low value
partial filled
pipe detection
Response time
part. filled
pipe detect.
End of wizard
A0013801-EN
å
21
"Partial filled pipe detection" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Description
Selection/
User entry
Factory setting
Assign process variable
Select a process variable to
detect empty or partially
filled pipes.
• Off
• Density
• Reference density
Density
Low value partial filled pipe
detection
Enter a lower limit value to
activate detection of an
empty or partially filled pipe.
Max. 15-digit, positive
floating-point number
Country-dependent:
• 0.2 kg/l
• 12.5 lb/cf
High value partial filled pipe
detection
Enter an upper limit value to
activate detection of an
empty or partially filled pipe.
Max. 15-digit, positive
floating-point number
Country-dependent:
• 6 kg/l
• 374.6 lb/cf
Response time part. filled
pipe detect.
Enter the time interval until
the diagnostic message
S862 Partly filled pipe
detection
is displayed for an
empty or partially filled pipe.
0 to 100 s
1 s