System integration
Proline Promass F 200 HART
50
Hauser
The assignment of the measured variables to the dynamic variables can be modified and assigned as
desired via local operation and the operating tool using the following parameters:
• Expert
®
Communication
®
HART output
®
Assign PV
• Expert
®
Communication
®
HART output
®
Assign SV
• Expert
®
Communication
®
HART output
®
Assign TV
• Expert
®
Communication
®
HART output
®
Assign QV
The following measured variables can be assigned to the dynamic variables:
Measured variables for PV (primary dynamic variable)
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
Measured variables for SV, TV, QV (secondary, tertiary and quaternary dynamic variable)
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
9.3
Other settings
In the
Configuration
submenu, you can configure other settings for the HART protocol (e.g. Burst
mode)
Navigation path
"Expert" menu
®
Communication
®
HART output
®
Configuration