Proline Promass F 200 HART
System integration
Hauser
49
9
System integration
9.1
Overview of device description files
9.1.1
Current version data for the device
Firmware version
01.02.zz
• On the title page of the Operating instructions
• On transmitter nameplate (
• Parameter
firmware version
Diagnostics
®
Device info
®
Firmware version
Release date of firmware version
07.2012
---
Manufacturer ID
0x11
Manufacturer ID
parameter
Diagnostics
®
Device info
®
Manufacturer ID
Device type ID
0x54
Device type
parameter
Diagnostics
®
Device info
®
Device type
HART protocol revision
6.0
---
Device revision
3
•
Device revision
parameter
Diagnostics
®
Device info
®
Device revision
For an overview of the different firmware versions for the device (
9.1.2
Operating tools
The following is a list of the suitable device description file for each individual operating tool with
source.
Operating tool via HART protocol
Sources for obtaining device descriptions
Field Xpert SFX100
Use update function of handheld terminal
FieldCare
•
®
Download Area
• CD–ROM (contact Hauser)
• DVD (contact Hauser)
AMS Device Manager
(Emerson Process Management)
®
Download Area
SIMATIC PDM
(Siemens)
®
Download Area
Field Communicator 475
(Emerson Process Management)
Use update function of handheld terminal
9.2
Measured variables via HART protocol
The following measured variables (HART device variables) are assigned to the dynamic variables at
the factory:
Dynamic variables
Measured variables
(HART device variables)
Primary dynamic variable (PV)
Mass flow
Secondary dynamic variable (SV)
Totalizer 1
Tertiary dynamic variable (TV)
Density
Quaternary dynamic variable (QV)
Temperature